Infinity Internal Hex System 204711Recalls & reported problems
Ace Surgical Supply CO., Inc. · infinity INTERNAL 4.7x11.5MM Ø4.5HD HEX IMPLANT 6AL-4V ELI · Class 2 · Implant
Quick summary
This ACE Surgical Infinity internal hex implant is a small screw-shaped dental implant used as a root-form anchor inside the jawbone. The 63 reports mostly describe the implant failing to fuse with the surrounding bone and becoming loose or lost, sometimes with pain, swelling, infection, or a fistula. All 63 reports are injuries; no deaths were reported. No recalls name this device, and no new reports have come in during the past year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 63
- Deaths reported
- 0
- Injuries reported
- 63
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 00614950004751), received between Mar 15, 2016 and Jan 6, 2024 — classified as 63 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (53 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem53
- Failure to Osseointegrate47
- Insufficient Device Problem Information7
- Patient Device Interaction Problem4
- Loss of Osseointegration1
Patient problems
- Failure of Implant63
- Pain8
- Unspecified Infection5
- Swelling2
- Fistula1
- Numbness1
Recent adverse event reports
The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.
InjuryJan 6, 2024Failure to Osseointegrate; Failure to Osseointegrate
InjuryAug 1, 2023Loss of Osseointegration; Loss of Osseointegration
InjuryJul 5, 2023Failure to Osseointegrate; Osseointegration Problem; Patient Device Interaction Problem; Failure to Osseointegrate; Osseointegration Problem; Patient Device Interaction Problem
InjuryMay 26, 2023Failure to Osseointegrate; Osseointegration Problem; Patient Device Interaction Problem; Failure to Osseointegrate; Osseointegration Problem; Patient Device Interaction Problem
InjuryOct 18, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12648034 ↗InjuryOct 18, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12647635 ↗InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, FISTULA AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER. DOCTOR IMPLANTED EXPIRED PRODUCT INTO PATIENT.
Patient problems: Failure of Implant; Fistula
MAUDE report 12567711 ↗InjuryAug 6, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12290898 ↗InjuryJun 21, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
WRONG SIZE IMPLANTED WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12035178 ↗InjuryMay 28, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
BONE QUALITY WAS TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED AT SURGERY. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11904048 ↗InjuryApr 22, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11711820 ↗InjuryMar 3, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, PAIN, BLEEDING, INFECTION AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT UNDERGOING CHEMOTHERAPY.
Patient problems: Failure of Implant; Pain; Unspecified Infection
MAUDE report 11404028 ↗InjuryMar 3, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, PAIN, BLEEDING, INFECTION AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS UNDERGOING CHEMOTHERAPY.
Patient problems: Failure of Implant; Pain
MAUDE report 11403998 ↗InjuryMar 3, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11403994 ↗InjuryMar 3, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, FISTULA AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 11403989 ↗InjuryFeb 2, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 11265741 ↗InjuryJan 14, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 11171094 ↗InjuryJan 7, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11133537 ↗InjuryDec 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INFECTION AND ASYMPTOMATIC WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED., AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 10936895 ↗InjuryNov 8, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10806319 ↗InjuryNov 8, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10806305 ↗InjuryNov 3, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10777746 ↗InjuryJul 21, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND BLEEDING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 10301196 ↗InjuryJun 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY, PAIN AND BLEEDING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS ALLERGIES AND HAD RADIATION TX-HEAD/NECK AREA.
Patient problems: Failure of Implant; Pain
MAUDE report 10108593 ↗InjuryApr 24, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED.
Patient problems: Failure of Implant
MAUDE report 9995495 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.