Infinity Internal Hex System 205108Recalls & reported problems
Ace Surgical Supply CO., Inc. · infinity INTERNAL 5.1x8.0MM Ø4.5HD HEX IMPLANT 6AL-4V ELI · Class 2 · Implant
All models of Infinity Internal Hex System (24)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is an ACE Surgical Infinity internal hex endosseous dental implant, a small screw-shaped fixture placed in the jaw to anchor a replacement tooth. The reports mostly describe the implant failing to fuse with the surrounding bone, leaving it loose or spinning, along with pain, numbness, swelling, and some infection or fistula. There are 54 reports, all describing injuries, and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 54
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00614950004799), received between Dec 28, 2015 and Jun 9, 2026 — classified as 54 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (44 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem44
- Failure to Osseointegrate37
- Insufficient Device Problem Information4
- Loss of Osseointegration2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Failure of Implant52
- Pain11
- Unspecified Infection6
- Numbness4
- Fistula3
- Swelling2
- Insufficient Information1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2026Failure to Osseointegrate
InjuryOct 12, 2023Failure to Osseointegrate; Failure to Osseointegrate
InjuryAug 11, 2023Loss of Osseointegration; Loss of Osseointegration
InjuryJul 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
InjuryJun 22, 2023Loss of Osseointegration; Loss of Osseointegration
AFTER 3 WEEKS IMPLANT JUST SPINNED AND SLIGHT INFECTION #30.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 17187163 ↗InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, NUMBNESS AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Numbness
MAUDE report 12567671 ↗InjurySep 2, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12417834 ↗InjurySep 2, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (ALLOGRAFT) WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12417828 ↗InjuryAug 6, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MOBILITY AND PERI-IMPLANTITIS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE (ALLOGRAFT) WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12290902 ↗InjuryMar 29, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INFECTION, INFLAMMATION AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 11581041 ↗InjuryMar 29, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INFECTION, PAIN, INCREASED SENSITIVITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER AND ALLERGIC TO PENICILLIN.
Patient problems: Failure of Implant; Pain; Unspecified Infection
MAUDE report 11581038 ↗InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 11423185 ↗InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUALITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11423178 ↗InjuryMar 3, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11404045 ↗InjuryJan 14, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION PROCEDURE WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 11171377 ↗InjuryJul 21, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND BLEEDING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 10301206 ↗InjuryMay 7, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10032715 ↗InjuryMay 7, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10032710 ↗InjuryApr 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 9916866 ↗InjuryApr 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY, NUMBNESS, FISTULA AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Numbness
MAUDE report 9916861 ↗InjuryFeb 24, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (ALLOGRAFT) WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 9743934 ↗InjuryJan 31, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 9655300 ↗InjuryNov 18, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 9334175 ↗InjuryOct 25, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY, BLEEDING AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 9241275 ↗InjuryJul 16, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN, LACK OF OSSEOINTEGRATION AND BLEEDING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 8792494 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page