Infinity Internal Hex System 205111Recalls & reported problems
Ace Surgical Supply CO., Inc. · infinity INTERNAL 5.1x11.5MM Ø4.5HD HEX IMPLANT 6AL-4V ELI · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This is an ACE Surgical Infinity Internal Hex endosseous (dental) implant. The 36 reports mostly describe implants that never fused with the surrounding bone, leaving them mobile and requiring removal, along with pain, swelling, infection and numbness. There were no deaths and 36 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 36
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 00614950004812), received between Nov 24, 2015 and Oct 12, 2023 — classified as 36 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem30
- Failure to Osseointegrate24
- Insufficient Device Problem Information2
- Loss of Osseointegration1
Patient problems
- Failure of Implant36
- Pain9
- Swelling6
- Unspecified Infection6
- Numbness3
- Fistula2
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
InjuryOct 12, 2023Failure to Osseointegrate; Failure to Osseointegrate
InjuryOct 12, 2023Failure to Osseointegrate; Failure to Osseointegrate
InjuryJul 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
InjuryJun 21, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (ALLOGRAFT) WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12035281 ↗InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11423197 ↗InjuryNov 8, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10806316 ↗InjuryFeb 20, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 9733029 ↗InjuryFeb 20, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
SWELLING, INFECTION AND PAIN WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (NUOSS) WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain; Swelling; Unspecified Infection
MAUDE report 9733025 ↗InjuryNov 15, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 9330696 ↗InjuryMay 6, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8584247 ↗InjuryApr 18, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND SWELLING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8527747 ↗InjuryFeb 18, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8347432 ↗InjuryFeb 18, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS HIGH BLOOD PRESSURE, OSTEOPOROSIS AND SINUS PRESSURE.
Patient problems: Failure of Implant
MAUDE report 8347426 ↗InjuryJan 25, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 8278825 ↗InjuryJan 2, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 8210470 ↗InjuryDec 14, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 8163830 ↗InjurySep 28, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant; Pain
MAUDE report 7919564 ↗InjuryJun 1, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
MOBILITY, FISTULA AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Fistula
MAUDE report 7560621 ↗InjuryMay 10, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 7504237 ↗InjuryDec 20, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 7135272 ↗InjurySep 7, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, NUMBNESS, PAIN AND INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Numbness; Pain
MAUDE report 6847746 ↗InjurySep 6, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 6844931 ↗InjurySep 6, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT TIME OF SURGERY. PATIENT HAS DIABETES.
Patient problems: Failure of Implant
MAUDE report 6844925 ↗InjuryApr 10, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
MOBILITY, INFECTION AND INADEQUATE BONE QUALITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. DOCTOR ALSO REPORTED IMPLANT LOST INTEGRATION.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 6477149 ↗InjuryApr 10, 2017Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem
MOBILITY AND INADEQUATE BONE QUALITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. IMPLANT LOST INTEGRATION ACCORDING TO THE DOCTOR.
Patient problems: Failure of Implant
MAUDE report 6477115 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page