TRICAM Dental Implant System 303508Recalls & reported problems
Ace Surgical Supply CO., Inc. · TRICAM 3.5X8MM 3.5HD INTERNAL IMPLANT 6Al-4V ELI · Class 2 · Implant
All models of TRICAM Dental Implant System (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is an ACE Surgical Supply TRICAM internal dental implant, a small titanium post used to anchor a replacement tooth in the jaw. The 55 reports mostly describe the implant failing to fuse with the surrounding bone, so it becomes loose or falls out, sometimes with pain, swelling, bleeding, infection or a fistula. All 55 reports involve injuries and no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 55
- Deaths reported
- 0
- Injuries reported
- 55
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 00614950005161), received between Dec 21, 2015 and Sep 30, 2024 — classified as 55 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (39 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem39
- Failure to Osseointegrate30
- Insufficient Device Problem Information7
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Fracture1
- Loss of Osseointegration1
- Patient Device Interaction Problem1
Patient problems
- Failure of Implant52
- Pain14
- Unspecified Infection7
- Fistula6
- Swelling5
- Numbness4
- Swelling/ Edema1
Recent adverse event reports
The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT FELL OUT OF THE MOUTH AFTER 2 WEEKS.
Patient problems: Failure of Implant
MAUDE report 20334070 ↗InjurySep 30, 2024Failure to Osseointegrate; Failure to Osseointegrate
InjuryFeb 7, 2024Failure to Osseointegrate; Patient-Device Incompatibility; Failure to Osseointegrate; Patient-Device Incompatibility
InjuryMay 10, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
InjuryOct 19, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MOBILITY, PAIN, FISTULA, BLEEDING, AND SWELLING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Fistula; Pain
MAUDE report 12657207 ↗InjuryJul 30, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
ASYMPTOMATIC AND SINUS PERFORATION WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS (NANO BONE) PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12254141 ↗InjuryApr 12, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, PAIN, BLEEDING, INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS DIABETES AND A THYROID PROBLEM.
Patient problems: Failure of Implant; Pain
MAUDE report 11648046 ↗InjuryJan 8, 2021Osseointegration Problem; Osseointegration Problem
MOBILITY, PAIN, BLEEDING, SWELLING AND FISTULA WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Fistula; Pain; Swelling/ Edema
MAUDE report 11141376 ↗InjuryJan 8, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
ASINTOMATICA WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11141327 ↗InjuryJan 8, 2021Osseointegration Problem; Osseointegration Problem
MOBILITY, PAIN AND BLEEDING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 11141323 ↗InjuryNov 16, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 10842701 ↗InjuryNov 16, 2020Unintended Movement; Unintended Movement
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 10842692 ↗InjuryNov 16, 2020Adverse Event Without Identified Device or Use Problem
PERI-IMPLANTATIS AND FISTULA WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10842609 ↗InjuryOct 28, 2020Loss of Osseointegration; Loss of Osseointegration
BONE LOSS WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10750888 ↗InjuryApr 24, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. PATIENT HAS DIABETES.
Patient problems: Failure of Implant
MAUDE report 9995528 ↗InjuryApr 24, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, INADEQUATE BONE QUALITY/ BONE QUANTITY, PAIN, SWELLING, NUMBNESS, BLEEDING, FISTULA AND ASYMPTOMATIC WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Fistula; Numbness; Pain
MAUDE report 9995502 ↗InjuryJan 3, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. MOBILITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 9543657 ↗InjuryNov 15, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
SINUS PERFORATION, MOBILITY AND PAIN WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS BLOOD COAGULATION DISORDER.
Patient problems: Failure of Implant; Pain
MAUDE report 9330666 ↗InjuryOct 4, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 9155312 ↗InjuryOct 3, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INFECTION, SEVERE BONE LOSS AROUND IMPLANT AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 9151106 ↗InjuryAug 1, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8849821 ↗InjuryAug 1, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND PERI-IMPLANTITIS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8849815 ↗InjuryJul 16, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
DOCTOR REPORTED COULD NOT ACHIEVE PRIMARY STABILITY. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETICE RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 8792491 ↗InjuryApr 10, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8498760 ↗InjuryJan 3, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
PAIN, PERI-IMPLANTITIS AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED.
Patient problems: Failure of Implant
MAUDE report 8211757 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026