TRICAM Dental Implant System 304313Recalls & reported problems
Ace Surgical Supply CO., Inc. · TRICAM 4.3X13MM 4.3HD INTERNAL IMPLANT 6Al-4V ELI NOBEL CLONE · Class 2 · Implant
All models of TRICAM Dental Implant System (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Ace Surgical TRICAM implant is placed into bone. Most of the 80 reports describe the implant not bonding with bone, losing bone around it and becoming loose. The reports include 80 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 80
- Deaths reported
- 0
- Injuries reported
- 80
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 80 adverse event reports that list this exact device (DI 00614950005246), received between Dec 1, 2015 and Jun 8, 2024 — classified as 80 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (71 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem71
- Failure to Osseointegrate52
- Insufficient Device Problem Information4
- Patient Device Interaction Problem2
- Inadequacy of Device Shape and/or Size1
- Loss of Osseointegration1
Patient problems
- Failure of Implant77
- Pain12
- Unspecified Infection5
- Swelling4
- Patient Problem/Medical Problem2
- Numbness1
Recent adverse event reports
The 25 most recent of 80 reports, in the reporter's own words as published by the FDA.
InjuryJun 8, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT PROGRESSIVELY LOST BONE AND BECAME MOBILE.
Patient problems: Failure of Implant
MAUDE report 19494826 ↗InjuryJun 8, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT PROGRESSIVELY LOST BONE AND BECAME MOBILE.
Patient problems: Failure of Implant
MAUDE report 19494825 ↗InjuryMar 28, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IMPLANT DID NOT THREAD IN BONE. CHANGED TO A LARGER SIZE.
Patient problems: Failure of Implant
MAUDE report 19000186 ↗InjurySep 26, 2023Loss of Osseointegration; Loss of Osseointegration
PATIENT HAD IMPLANT RETAINED DENTURE - RECENTLY DEVELOPED MOBILITY & IMPLANT WAS REMOVED.
Patient problems: Failure of Implant
MAUDE report 17822025 ↗InjurySep 10, 2023Osseointegration Problem; Osseointegration Problem
InjuryJun 26, 2023Failure to Osseointegrate; Osseointegration Problem; Patient Device Interaction Problem; Failure to Osseointegrate; Osseointegration Problem; Patient Device Interaction Problem
InjuryJun 21, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
ASYMPTOMATIC WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 12035192 ↗InjuryMay 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
ASINTOMATICA WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11777060 ↗InjuryApr 22, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
ASYMPTOMATIC WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER AND HAS DIABETES.
Patient problems: Failure of Implant
MAUDE report 11711826 ↗InjuryApr 12, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 11648044 ↗InjuryApr 12, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 11648040 ↗InjuryNov 16, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10842610 ↗InjuryMay 7, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10032706 ↗InjuryApr 24, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
NUMBNESS WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant; Numbness
MAUDE report 9995518 ↗InjuryNov 18, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 9335320 ↗InjuryNov 18, 2019Osseointegration Problem; Patient Device Interaction Problem; Osseointegration Problem; Patient Device Interaction Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 9335317 ↗InjuryOct 24, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. MOBILITY AND PERI-IMPLANTITIS WERE REPORTED. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 9232371 ↗InjuryOct 24, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. MOBILITY AND PAIN WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Pain
MAUDE report 9232362 ↗InjuryOct 4, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY AND ASYMPTOMATIC WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (L-PRF) WAS PERFORMED. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 9155305 ↗InjuryOct 3, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. INADEQUATE BONE QUALITY/BONE QUANTITY, PAIN, PERI-IMPLANTITIS AND POOR ORAL HYGIENE WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS A PSYCHOLOGICAL DISORDER.
Patient problems: Failure of Implant; Pain
MAUDE report 9151107 ↗InjuryAug 1, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 8849817 ↗InjuryAug 1, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8849814 ↗InjuryAug 1, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8849813 ↗InjuryJul 16, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (BIO-OSS) WAS PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 8792492 ↗InjuryJun 27, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 8738611 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.