TRICAM Dental Implant System 305011Recalls & reported problems

Ace Surgical Supply CO., Inc. · TRICAM 5.0X11.5MM 5HD INTERNAL IMPLANT 6Al-4V ELI NOBEL CLONE · Class 2 · Implant

All models of TRICAM Dental Implant System (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Tricam root-form dental implant made by Ace Surgical Supply. The 27 reports mostly describe the implant not bonding with the surrounding bone, so it becomes loose and has to be removed, with reports also mentioning pain, swelling, infection, a fistula, and numbness. All 27 reports are injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 00614950005284), received between Feb 10, 2016 and May 18, 2026 — classified as 27 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem19
  • Failure to Osseointegrate17
  • Insufficient Device Problem Information3
  • Loss of Osseointegration2
  • Device Issue1

Patient problems

  • Failure of Implant26
  • Pain8
  • Swelling5
  • Unspecified Infection3
  • Fistula2
  • Numbness2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 18, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 25209286 ↗
  • InjuryFeb 7, 2024Failure to Osseointegrate; Failure to Osseointegrate

    DID NOT OSSEOINTEGRATE AFTER 8 WEEK HEALING PERIOD.

    Patient problems: Failure of Implant

    MAUDE report 18659547 ↗
  • InjuryAug 17, 2023Failure to Osseointegrate; Failure to Osseointegrate

    NO PRIMARY STABILITY.

    Patient problems: Failure of Implant

    MAUDE report 17568630 ↗
  • InjuryAug 16, 2023Loss of Osseointegration; Loss of Osseointegration

    NO PRIMARY STABILITY.

    Patient problems: Failure of Implant

    MAUDE report 17558507 ↗
  • InjuryJun 17, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY, INCREASED SENSITIVITY, INFLAMMATION AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant

    MAUDE report 12020293 ↗
  • InjuryMar 29, 2021Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    MOBILITY AND BONE LOSS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (ALLOGRAFT) WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11581043 ↗
  • InjuryJan 8, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11141322 ↗
  • InjuryJan 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    ASYMPTOMATIC WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11120766 ↗
  • InjuryMay 7, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND SWELLING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Swelling

    MAUDE report 10032720 ↗
  • InjuryApr 24, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/ BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. OSSEOINTEGRATION WAS NOT ACHIEVED.

    Patient problems: Failure of Implant

    MAUDE report 9995512 ↗
  • InjuryNov 15, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 9330671 ↗
  • InjuryNov 15, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND PAIN WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain

    MAUDE report 9330668 ↗
  • InjuryAug 1, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    PAIN, BLEEDING, INFECTION AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain; Unspecified Infection

    MAUDE report 8849811 ↗
  • InjuryApr 10, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8498752 ↗
  • InjuryNov 28, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY/BONE "QUANITTY" WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8110148 ↗
  • InjuryJul 23, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    SWELLING AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE III. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED (FREEZE DRIED DEMINERALIZED BONE).

    Patient problems: Failure of Implant; Swelling

    MAUDE report 7710703 ↗
  • InjuryNov 22, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MOBILITY, PAIN AND BRUXISM WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain

    MAUDE report 7052805 ↗
  • InjuryNov 2, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    PAIN WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED.

    Patient problems: Failure of Implant; Pain

    MAUDE report 6996211 ↗
  • InjuryOct 16, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    PAIN, SWELLING, NUMBNESS, ASINTOMATICA, AND FISTULA WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT TIME OF SURGERY. INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED.

    Patient problems: Failure of Implant; Fistula; Pain; Swelling

    MAUDE report 6951951 ↗
  • InjurySep 6, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    PAIN, BLEEDING, SWELLING, NUMBNESS AND FISTULA WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.

    Patient problems: Failure of Implant; Numbness; Pain; Swelling

    MAUDE report 6845029 ↗
  • InjuryAug 30, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INFECTION, MOBILITY AND FISTULA WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.

    Patient problems: Failure of Implant; Fistula; Unspecified Infection

    MAUDE report 6829684 ↗
  • InjuryMar 23, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    PAIN, SWELLING, BLEEDING, INFECTION AND NUMBNESS WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I.

    Patient problems: Failure of Implant; Numbness; Pain; Swelling; Unspecified Infection

    MAUDE report 6429868 ↗
  • InjuryDec 14, 2016Osseointegration Problem; Osseointegration Problem

    NO OSSEOINTEGRATION OF THE IMPLANT AFTER DISCOVERY BUT BEFORE FINAL RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT SURGERY TYPE IV.

    Patient problems: Failure of Implant

    MAUDE report 6173281 ↗
  • InjuryJun 22, 2016Device Issue; Device Issue

    PAIN WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE BETWEEN TYPE II & TYPE III. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PATIENT REQUESTED REMOVAL OF IMPLANT DUE TO PAIN.

    Patient problems: Failure of Implant; Pain

    MAUDE report 5741283 ↗
  • InjuryJun 2, 2016Osseointegration Problem; Osseointegration Problem

    MOBILITY WAS REPORTED. NO OSSEOINTEGRATION OF THE IMPLANT 1-3 MONTHS AFTER FINAL RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II.

    Patient problems: Failure of Implant

    MAUDE report 5695370 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →