TRICAM Dental Implant System 305013Recalls & reported problems

Ace Surgical Supply CO., Inc. · TRICAM 5.0X13MM 5HD INTERNAL IMPLANT 6Al-4V ELI NOBEL CLONE · Class 2 · Implant

All models of TRICAM Dental Implant System (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This TRICAM dental implant is a root-form implant placed in the jaw to hold dental work. Reports mostly describe implants that did not join to the bone and became loose, often removed before the crown was placed, with some mentioning pain or infection. There are 23 reports naming its identifier, with 23 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
23
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00614950005291), received between Dec 22, 2015 and Feb 7, 2024 — classified as 23 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem17
  • Failure to Osseointegrate15
  • Insufficient Device Problem Information3
  • Loss of Osseointegration1
  • Patient-Device Incompatibility1

Patient problems

  • Failure of Implant23
  • Pain3
  • Unspecified Infection3
  • Swelling1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 7, 2024Failure to Osseointegrate; Patient-Device Incompatibility; Failure to Osseointegrate; Patient-Device Incompatibility

    SAME DAY SPINNER.

    Patient problems: Failure of Implant

    MAUDE report 18664289 ↗
  • InjuryOct 5, 2023Loss of Osseointegration; Loss of Osseointegration

    PT HAD IMPLANT RETAINED DENTURE - RECENTLY DEVELOPED MOBILITY - IMPLANT REMOVED.

    Patient problems: Failure of Implant

    MAUDE report 17882425 ↗
  • InjuryJun 21, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12035199 ↗
  • InjuryApr 12, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    ASYMPTOMATIC WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WA ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (VITOSS) WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11648037 ↗
  • InjuryFeb 25, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11377564 ↗
  • InjuryFeb 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    ASYMPTOMATIC, POOR ORAL HYGIENE AND INADEQUATE BONE QUALITY/BONE QUANTITY WEE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED AT THE TIME OF SURGERY. PATIENT HAS ALLERGIES (PCN).

    Patient problems: Failure of Implant

    MAUDE report 11282507 ↗
  • InjuryNov 16, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 10842654 ↗
  • InjuryOct 24, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. MOBILITY AND ASYMPTOMATIC WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 9232358 ↗
  • InjuryAug 1, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND PERI-IMPLANTITIS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8849812 ↗
  • InjuryJun 27, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8738618 ↗
  • InjuryApr 17, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MOBILITY, PERI-IMPLANTITIS AND PAIN WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE I. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE (BONE GRAFT) WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain

    MAUDE report 7436145 ↗
  • InjuryFeb 26, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND ASYMPTOMATIC WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 7295598 ↗
  • InjuryDec 20, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MOBILITY, INFECTION, PERI-IMPLANTITIS, ASINTOMATICA AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS XEROSTOMIA.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 7135225 ↗
  • InjuryNov 13, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.

    Patient problems: Failure of Implant

    MAUDE report 7024649 ↗
  • InjuryNov 13, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.

    Patient problems: Failure of Implant

    MAUDE report 7024647 ↗
  • InjurySep 5, 2017Osseointegration Problem; Osseointegration Problem

    ASYMPTOMATIC WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. PATIENT HAS DIABETES MELLITUS.

    Patient problems: Failure of Implant

    MAUDE report 6843458 ↗
  • InjuryApr 24, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    OSSEOINTEGRATION WAS NOT ACHIEVED. BONE QUALITY AT TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION.

    Patient problems: Failure of Implant

    MAUDE report 6516595 ↗
  • InjuryApr 24, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    OSSEOINTEGRATION WAS NOT ACHIEVED. BONE QUALITY AT TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION.

    Patient problems: Failure of Implant

    MAUDE report 6516563 ↗
  • InjuryMar 24, 2017Osseointegration Problem; Osseointegration Problem

    PAIN, BLEEDING, SWELLING, INFECTION AND NUMBNESS WERE REPORTED. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. PATIENT HAS A BLOOD CLOTTING DISORDER.

    Patient problems: Failure of Implant; Pain; Swelling; Unspecified Infection

    MAUDE report 6432564 ↗
  • InjuryFeb 27, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MOBILITY AND INFECTION WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION WAS PERFORMED AT TIME OF SURGERY.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 6360674 ↗
  • InjuryOct 28, 2016Osseointegration Problem; Osseointegration Problem

    MOBILITY AND PAIN WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II.

    Patient problems: Failure of Implant; Pain

    MAUDE report 6063940 ↗
  • InjuryApr 8, 2016Osseointegration Problem; Osseointegration Problem

    ASYMPTOMATIC WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE RESTORATION. PRIMARY STABILITY WAS ACHIEVED BUT OSSEOINTEGRATION WAS NOT. AUGMENTATION PROCEDURE WAS PERFORMED AT THE TIME OF SURGERY. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE II. IT WAS REPORTED IMPLANT CAME OUT WHEN WENT TO UNCOVER.

    Patient problems: Failure of Implant

    MAUDE report 5564120 ↗
  • InjuryDec 22, 2015Failure to Osseointegrate

    MOBILITY WAS REPORTED AT THE TIME OF IMPLANT FAILURE.

    Patient problems: Failure of Implant

    MAUDE report 5319904 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →