Infinity External Hex Implant System 403711Recalls & reported problems
Ace Surgical Supply CO., Inc. · INFINITY EXTERNAL 3.75X11.5MM Ø4.0HD HEX IMPLANT TI CP GRADE 4 · Class 2 · Implant
All models of Infinity External Hex Implant System (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This ACE Surgical Infinity external hex dental implant is a screw-shaped titanium fixture placed in the jawbone to anchor a tooth replacement. Across 44 reports, all describe injuries with no deaths, and most describe implants that never fused properly with the bone or later lost that fusion, leading to pain, swelling, numbness, infection around the implant, fistula, and a loose or failed implant. The reports run from 2015 to 2024, with none in the past 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 44
- Deaths reported
- 0
- Injuries reported
- 44
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00614950005420), received between Dec 1, 2015 and Jun 4, 2024 — classified as 44 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (35 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem35
- Failure to Osseointegrate14
- Insufficient Device Problem Information7
- Loss of Osseointegration1
Patient problems
- Failure of Implant42
- Pain12
- Unspecified Infection10
- Numbness5
- Swelling5
- Fistula4
- Patient Problem/Medical Problem1
Recent adverse event reports
The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.
InjuryJun 4, 2024Loss of Osseointegration; Loss of Osseointegration
InjuryMay 24, 2023Osseointegration Problem; Osseointegration Problem
TIME OF LOSS IN RELATION TO THE IMPLEMENTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION.
Patient problems: Failure of Implant
MAUDE report 16992352 ↗InjuryMar 29, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 11581164 ↗InjuryNov 16, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10842694 ↗InjuryNov 16, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 10842693 ↗InjuryJun 27, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INADEQUATE BONE QUALITY/BONE QUANTITY, PAIN, SWELLING, BLEEDING, NUMBNESS, MOBILITY AND FISTULA WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Fistula; Numbness; Swelling
MAUDE report 8738607 ↗InjuryMar 19, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
PAIN WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant; Pain
MAUDE report 8431368 ↗InjuryOct 19, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, PERI-IMPLANTITIS AND ASYMPTOMATIC WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant
MAUDE report 7983349 ↗InjuryOct 19, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, PAIN AND POOR ORAL HYGIENE WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1-5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 7983329 ↗InjuryMay 10, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY, SWELLING, POOR ORAL HYGIENE AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Pain; Swelling
MAUDE report 7504241 ↗InjuryMay 7, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
INFECTION, ASINTOMATICA, BAD ORAL HYGIENE AND PERI-IMPLANTITIS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 7490104 ↗InjuryJan 9, 2018Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEONTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant
MAUDE report 7174653 ↗InjuryDec 21, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant
MAUDE report 7137428 ↗InjuryNov 14, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
MOBILITY AND PAIN WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Pain
MAUDE report 7027519 ↗InjuryNov 13, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 7026000 ↗InjuryNov 13, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
INFECTION, BLEEDING, SWELLING AND NUMBNESS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED.
Patient problems: Failure of Implant; Numbness; Swelling; Unspecified Infection
MAUDE report 7025989 ↗InjuryNov 13, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
PAIN, FISTULA AND INFECTION WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED.
Patient problems: Failure of Implant; Fistula; Pain; Unspecified Infection
MAUDE report 7024616 ↗InjuryNov 13, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
INADEQUATE BONE QUALITY/BONE QUANTITY, MOBILITY, INFECTION, PAIN, BLEEDING, SWELLING, NUMBNESS, PERI-IMPLANTITIS AND FISTULA WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. PATIENT IS A SMOKER.
Patient problems: Failure of Implant; Numbness; Pain; Unspecified Infection
MAUDE report 7024614 ↗InjurySep 5, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BLEEDING, SWELLING, NUMBNESS, FISTULA AND PERI-IMPLANTITIS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. PATIENT IS A SMOKER AND HAS XEROSTOMIA.
Patient problems: Failure of Implant; Fistula; Numbness; Swelling
MAUDE report 6841396 ↗InjurySep 5, 2017Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem
BLEEDING, SWELLING, NUMBNESS, FISTULA AND PERI-IMPLANTITIS WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. PATIENT IS A SMOKER AND HAS XEROSTOMIA.
Patient problems: Failure of Implant; Fistula; Numbness; Swelling
MAUDE report 6841393 ↗InjuryApr 24, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
BLEEDING AND MOBILITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1-5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PATIENT HAS PERI-IMPLANTITIS.
Patient problems: Failure of Implant
MAUDE report 6517035 ↗InjuryFeb 23, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
PAIN WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1-5 YRS AFTER FINAL PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. IT WAS ALSO REPORTED FRACTURE OF THE IMPLANT.
Patient problems: Failure of Implant
MAUDE report 6353367 ↗InjuryDec 29, 2016Osseointegration Problem; Osseointegration Problem
INFECTION WAS REPORTED. NO OSSEOINTEGRATION 1-3 MONTHS BEFORE RESTORATION.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 6214365 ↗InjuryNov 7, 2016Osseointegration Problem; Osseointegration Problem
MOBILITY AND INFECTION WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS OVER A YEAR AFTER FINAL PROSTHETIC RESTORATION. IT WAS ALSO REPORTED NO OSSEOINTEGRATION OF THE IMPLANT.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 6083432 ↗InjuryNov 7, 2016Osseointegration Problem; Osseointegration Problem
PAIN WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS MORE THAN A YEAR AFTER PROSTHETIC RESTORATION. IT WAS ALSO REPORTED BREAKAGE OF THE IMPLANT.
Patient problems: Failure of Implant; Pain
MAUDE report 6083387 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page