ALLEN 107003 (REF A-71101-UK)Recalls & reported problems
Hill-Rom, Inc. · Allen Advance Table · Class 1
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00615521003104), received between May 28, 2021 and Aug 20, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure1
- Detachment of Device or Device Component1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 20, 2025Unintended Movement; Unintended Movement
BAXTER RECEIVED A REPORT STATING THAT ALLEN ADVANCE TABLE WAS GOING UP/DOWN ON ITS OWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22853548 ↗MalfunctionOct 15, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
HILLROM RECEIVED A USER REPORT, STATING THAT WHILE POSITIONING A PATIENT PRONE FOR A SPINAL PROCEDURE THE HEAD SUPPORT ATTACHMENT CAME LOOSE FROM AN ALLEN ADVANCE SURGICAL TABLE PUTTING THE PATIENT AT RISK FOR UNSUPPORTED HEAD DROPPING TO OCCUR. THE CUSTOMER CONFIRMED THERE WAS NO REPORT OF INJURY TO THE PATIENT ASSOCIATED WITH THIS EVENT AS THE HEAD WAS STILL BEING HELD BY SURGICAL STAFF DURING POSITIONING OF THE PATIENT. THE CUSTOMER NOTES AFTER POSITIONING THE PATIENT THE ¿PRONEVIEW¿ PASSED VISUAL INSPECTION BY THE SURGICAL TEAM AND ¿SEEMED STURDY¿ FOR PATIENT USE. HOWEVER, FOLLOWING PATIENT REPOSITIONING, THE STARBURST ADAPTOR CONNECTING THE C-FLEX HEAD POSITIONER TO THE ALLEN TABLE CAME LOOSE. IT IS NOTED THE FACILITY¿S CLINICIAN REVIEW STATEMENT REPORTED THAT THE EVENT ¿MOST LIKELY THE EQUIPMENT WAS USED INCORRECTLY, AND TRAINING IS NEEDED¿. HRC TECHNICIAN INSPECTED THE ALLEN TABLE ON SITE AND THE CUSTOMER SENT IN THE C-FLEX HEAD POSITIONER FOR FURTHER INVESTIGATION. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4) FOR THE ALLEN ADVANCE TABLE A-71101-UK TO WHICH THE C-FLEX HEAD POSITIONER CONNECTS FOR WHICH WE ARE FILING THE EMDR AND THEREFORE WE WIL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12639732 ↗MalfunctionMay 28, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE THE ALLEN ADVANCE TABLE WOULD NOT MOVE, THE LIGHTS ON THE TOP COLUMN WERE NOT SHOWING, THERE WERE NO FUNCTIONS VIA THE PENDANT, AND THE BED REMAINED UNLOCKED. AT THE CUSTOMER'S REQUEST A HILLROM REPRESENTATIVE ATTENDED THE CUSTOMER SITE AND ATTEMPTED TO RESOLVE THE ISSUE, HOWEVER THEY WERE UNABLE TO SUCCESSFULLY GET THE DEVICE TO FUNCTION CORRECTLY. FOLLOWING THE UNSUCCESSFUL ATTEMPT TO REPAIR THE DEVICE, THE PATIENT WAS TRANSFERRED TO A BACKUP TABLE FOR COMPLETION OF THE SURGICAL PROCEDURE. THIS RESULTED IN A DELAY OF GREATER THAN 1 HOUR TO THE SURGICAL PROCEDURE. SUBSEQUENTLY, THE CUSTOMER CONFIRMED THAT THEY HAD EXPERIENCED A SIMILAR ISSUE WITH THE SAME DEVICE A WEEK PRIOR TO THIS INCIDENT BUT THAT THE ISSUE HAD BEEN SOLVED WHEN THE CUSTOMER PULLED THE EMERGENCY BRAKES AND TURNED THE TABLE OFF. THERE WAS NO REPORTED INJURY TO THE AFFECTED PATIENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11901628 ↗
Other Table, operating-room, electrical devices
- NA 72203299Smith & Nephew, Inc. · TABLE, OPERATING-ROOM, ELECTRICAL298 reports
- TS 7000 1723633Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall22 reports
- ALLEN 107003Hill-Rom, Inc. · Table, operating-room, electrical1 recall9 reports
- Head-Positioning accessories 1739994Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall1 report
- ALLEN 107003Hill-Rom, Inc. · Table, operating-room, electrical1 recall0 reports
- TS 7500 4091000Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 1717021Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 2067886Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.