MEDRAD® Stellant Sterile Disposable Syringe SSS-CTP-SPK (UDI 00616258005218)Recalls & reported problems
Bayer Medical Care Inc. · Stellant Single Syringe Kit with LPDCT with T-Connector and 2 sterile spikes (CT-STELLANT SINGLE SYR KIT, LPCT, SPIKE) (60728818) · Class 2
All models of MEDRAD Stellant Sterile Disposable Syringe (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00616258005218), received between Jul 12, 2017 and Mar 12, 2020 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (5 reports).
6 other recalls cover devices cleared under the same FDA submission (K023183); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Manufacturing, Packaging or Shipping Problem5
- Product Quality Problem3
- Contamination /Decontamination Problem2
- Particulates2
- Air Leak1
- Clumping in Device or Device Ingredient1
- Defective Component1
- Device Contamination with Chemical or Other Material1
Patient problems
- Not Applicable5
- No Known Impact Or Consequence To Patient3
- No Patient Involvement3
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 12, 2020Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING A STELLANT CT DISPOSABLE SYRINGE KIT A FOREIGN MATERIAL WAS VISUALIZED IN THE TUBING. THE CUSTOMER ELECTED NOT TO USE THE SYRINGE KIT. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 9824178 ↗MalfunctionAug 9, 2019Product Quality Problem; Manufacturing, Packaging or Shipping Problem; Product Quality Problem; Manufacturing, Packaging or Shipping Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT CT DISPOSABLE KIT THE CUSTOMER OBSERVED A FOREIGN BODY INSIDE THE TUBING. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: Not Applicable
MAUDE report 8878893 ↗MalfunctionFeb 26, 2019Product Quality Problem; Manufacturing, Packaging or Shipping Problem; Product Quality Problem; Manufacturing, Packaging or Shipping Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT CT DISPOSABLE KIT THE CUSTOMER OBSERVED A FOREIGN BODY INSIDE THE TUBING. THE CUSTOMER ELECTED NOT TO USE THE SYRINGE KIT. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: Not Applicable
MAUDE report 8370624 ↗MalfunctionOct 31, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT CT DISPOSABLE KIT THE CUSTOMER OBSERVED A BLACK PARTICULATE INSIDE THE TUBING. THE CUSTOMER ELECTED NOT TO USE THE SYRINGE KIT. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: Not Applicable
MAUDE report 8023550 ↗MalfunctionOct 24, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT CT DISPOSABLE KIT AND UPON INSPECTION, THEY SAW A FOREIGN BODY WITHIN THE LOW PRESSURE CONNECTOR TUBING (LPCT). THE CUSTOMER ELECTED NOT TO USE THE TUBING. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: Not Applicable
MAUDE report 8000425 ↗MalfunctionAug 3, 2018Device Contamination with Chemical or Other Material; Manufacturing, Packaging or Shipping Problem; Device Contamination with Chemical or Other Material; Manufacturing, Packaging or Shipping Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLENT CT DISPOSABLE KIT AND UPON INSPECTION, THEY SAW A FOREIGN BODY WITHIN THE LOW PRESSURE CONNECTOR TUBING (LPCT). THE CUSTOMER ELECTED NOT TO USE THE TUBING. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 7747557 ↗MalfunctionJul 19, 2018Product Quality Problem; Defective Component; Manufacturing, Packaging or Shipping Problem; Product Quality Problem; Defective Component; Manufacturing, Packaging or Shipping Problem
THE SITE REPORTED THE FOLLOWING: A FOREIGN BODY WAS FOUND INSIDE THE LOW PRESSURE CONNECTOR TUBING (LPCT) OF A SSS-CTP-SPK SYRINGE KIT. THE KIT WAS REMOVED FROM THE AREA AND WAS NOT USED FOR A PATIENT.
Patient problems: Not Applicable
MAUDE report 7704627 ↗MalfunctionMay 14, 2018Clumping in Device or Device Ingredient; Sediment, Precipitate Or Deposit In Device Or Device Ingredient; Clumping in Device or Device Ingredient; Sediment, Precipitate Or Deposit In Device Or Device Ingredient
THE SITE REPORTED THE FOLLOWING: A STELLANT DISPOSABLE SET HAS VISIBLE FOREIGN MATERIAL IN THE TUBING. THIS IS A SSS-CPT-SPK SYRINGE KIT. THE KIT WAS REMOVED FROM THE AREA AND WAS NOT USED FOR A PATIENT.
Patient problems: No Patient Involvement
MAUDE report 7510600 ↗MalfunctionJan 31, 2018Particulates
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT CT DISPOSABLE KIT THE CUSTOMER OBSERVED A BLACK PARTICULATE INSIDE THE TUBING. THE CUSTOMER ELECTED NOT TO USE THE SYRINGE KIT. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7230619 ↗MalfunctionJul 12, 2017Particulates
THE SITE REPORTED THE FOLLOWING: A RADIOLOGIC TECHNOLOGIST REMOVED A SYRINGE FROM THE SYRINGE KIT PACKAGING, SPIKED A BOTTLE OF CONTRAST MEDIA AND BEGAN TO FILL THE SYRINGE WITH CONTRAST MEDIA. ABOUT 2/3 OF THE CONTRAST WAS INTO THE SYRINGE WHEN HE NOTED 1 OR 2 BLACK FLECKS FLOATING IN THE CONTRAST. HE STOPPED FILLING THE SYRINGE AND TOOK A PHOTO OF THE BLACK FLECK. HE THEN REMOVED THE SYRINGE AND DISCARDED IT. THE SYRINGE WAS NEVER USED FOR A PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6706430 ↗MalfunctionJul 12, 2017Air Leak
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT CT DISPOSABLE KIT AND UPON INSPECTION, THEY SAW BUBBLES OF WATER WITHIN THE LOW PRESSURE CONNECTOR TUBING (LPCT). THE CUSTOMER ELECTED NOT TO USE THE TUBING. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6705813 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K023183. One submission often covers many models, so check the product and lot details in each record.
- Z-0745-2019 · Dec 14, 2018 · TerminatedMAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Inje
- Z-0147-2012 · Oct 24, 2011 · TerminatedMedrad(R) Sterile Disposable Stellant Syringe Kit, Catalog # SDS-CTP-SPK, Lot # 112865. Angiographic syringe intended to be used for the specific purpose of injecting intravenous
- Z-1954-2010 · May 24, 2010 · TerminatedStellant CT Dual Syringe Kits, Catalog Number SDS-CTP-QFT and SDS-CTP-SPK
- Z-0154-06 · Oct 10, 2005 · TerminatedStellant CT Injection System. Powered Injector with Syringe. Angiographic Injector and Syringe. Model numbers S, SX, D, and DX
- Z-0257-05 · Nov 3, 2004 · TerminatedStellant Dual Syringe Kits. CT Injector and Accessories. Catalog Number SDS-CTP-QFT and SDS-CTP-SPK.
- Z-0039-05 · Oct 8, 2004 · TerminatedStellant Syringe Kit. Sterile Disposable Syringe Kits. CT Injector and Accessories. Catalog number SSS-CTP-QFT
Other Injector And Syringe, Angiographic devices
- Monarch® 00884450298022Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls37 reports
- Blue Diamond™ 00884450498743Merit Medical Systems, Inc. · Injector and syringe, angiographic2 recalls0 reports
- Monarch® 00884450298053Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- Monarch® 00884450298060Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- NAMIC 80085007Medline Industries, Inc. · Injector and syringe, angiographic1 recall36 reports
- NA 622510B. Braun Medical Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC85 reports
- NAMIC 80087007Medline Industries, Inc. · Injector and syringe, angiographic1 recall20 reports
- Blue Diamond™ 00884450381168Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page