Skip to content

MEDRAD® Avanta Fluid Management Injection System (REF AVA 500 TABL, AVA 500 PEDL)Recalls & reported problems

Bayer Medical Care Inc. · "ASSEMBLY,INJ SYSTEM,MEDRAD AVANTA,MC (60764652)" · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
1
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00616258008578), received between Oct 9, 2018 and Nov 18, 2025 — classified as 1 death, 4 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

6 other recalls cover devices cleared under the same FDA submission (K050456); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Insufficient Device Problem Information2

Patient problems

  • Air Embolism5
  • Cardiac Arrest2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • DeathNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING INFORMATION WAS REPORTED TO BAYER. A PATIENT WAS UNDERGOING A CORONARY ANGIOGRAM WHILE CONNECTED TO A MEDRAD® AVANTA FLUID MANAGEMENT SYSTEM (SERIAL NUMBER (B)(6). DURING THE FIRST INJECTION OF CONTRAST, AIR WAS VISUALIZED ON THE DISPLAYED IMAGES AFTER WHICH NON-SPECIFIED EMERGENCY TREATMENT WAS IMMEDIATELY INITIATED, AND CARDIOPULMONARY RESUSCITATION (CPR) / ADVANCED CARDIOVASCULAR LIFE SUPPORT (ACLS) WAS PERFORMED BY THE MEDICAL TEAM. DESPITE THESE EFFORTS, THE PATIENT COULD NOT BE RESUSCITATED AND CEASED TO BREATHE.

    Patient problems: Air Embolism

    MAUDE report 23582317 ↗
  • InjuryJan 23, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    BAYER MEDICAL CARE INC. WAS INFORMED THAT A 60-YEAR-OLD MALE EXPERIENCED AN ALLEGED AIR INJECTION DURING A HEART CATHETERIZATION PROCEDURE WHILE CONNECTED TO A MEDRAD® AVANTA FLUID MANAGEMENT INJECTION SYSTEM (SN (B)(6)). THE CUSTOMER REPORTED THAT AN UNDISCLOSED AMOUNT OF AIR WAS VISUALIZED AFTER A CONTRAST INJECTION IN THE LEFT CORONARY SYSTEM. FOLLOWING THE EVENT, THE PATIENT WENT INTO CARDIAC ARREST, WAS IMMEDIATELY RESUSCITATED AND WAS SUCCESSFULLY STABILIZED. THE PATIENT IS REPORTED AS DOING WELL.

    Patient problems: Air Embolism; Cardiac Arrest

    MAUDE report 21218217 ↗
  • InjuryFeb 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED TO A BAYER REPRESENTATIVE THAT AN ALLEGED AIR INJECTION OCCURRED ON A (B)(6) MALE PATIENT WHILE UTILIZING THE AVANTA FLUID MANAGEMENT SYSTEM ((B)(4)) TO PERFORM A CORONARY ANGIOGRAPHY PROCEDURE. DURING THE FIRST INJECTION, WITH THE CATHETER LOCATED IN THE LEFT CORONARY ARTERY, 5 ML OF AIR WAS OBSERVED ON THE FLUOROSCOPIC IMAGES. THE PATIENT SUFFERED A TRANSIENT CARDIAC ARREST BUT WAS REPORTED TO LATER RECOVER.

    Patient problems: Air Embolism

    MAUDE report 13536776 ↗
  • InjurySep 11, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    ON (B)(6) 2019, THE CUSTOMER REPORTED THAT A PATIENT SUFFERED A CARDIAC ARREST AFTER AN UNDISCLOSED AMOUNT OF AIR WAS ALLEGEDLY INJECTED WHILE CONNECTED TO AN AVANTA FLUID MANAGEMENT SYSTEM. POST RESUSCITATION, A BALLOON PUMP WAS INSERTED AND THE PATIENT WAS TRANSFERRED TO THE INTENSIVE CARE UNIT (ICU) IN STABLE CONDITION FOR FURTHER EVALUATION AND CARE.

    Patient problems: Air Embolism; Cardiac Arrest

    MAUDE report 8989333 ↗
  • InjuryOct 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A BAYER REPRESENTATIVE REPORTED THE FOLLOWING: A (B)(6), MALE PATIENT WITH AN ADMITTING DIAGNOSIS OF CORONARY ARTERY DISEASE WAS CONNECTED TO THE AVANTA FLUID MANAGEMENT SYSTEM WHILE UNDERGOING A CARDIAC CATHETERIZATION PROCEDURE WITH PCI (PERCUTANEOUS CORONARY INTERVENTION). AN ALLEGED AIR INJECTION OCCURRED DURING THE FINAL IMAGE POST PCI WITHIN THE RIGHT CORONARY ARTERY (RCA) FOLLOWED BY A DISSECTION AT THE BIFURCATION OF THE RCA AND ANOMALOUS CIRCUMFLEX. A BAYER CLINICAL SUPPORT SPECIALIST WAS PRESENT FOR THE CASE AND PROVIDED THAT THE OPINION OF THE PHYSICIAN PERFORMING THE CASE WAS THE AIR WAS INTRODUCED VIA A SIDE ARM ADAPTOR THAT WAS ATTACHED BETWEEN THE SPAT AND THE PATIENT CATHETER. ADDITIONALLY, THE PHYSICIAN PROVIDED THAT THE MOST PROBABLE CAUSE OF THE DISSECTION WAS MOST LIKELY THE PATIENT CATHETER AND NOT THE INJECTOR; HE SURMISED AN AIR BUBBLE MAY CONTRIBUTE TO THE SIZE OF THE DISSECTION, BUT IS UNLIKELY THE CAUSE. THE PATIENT EXPERIENCED CHEST PAIN AND REQUIRED MEDICAL MANAGEMENT ALONG WITH ADMISSION TO THE STEP DOWN CARDIAC CARE UNIT. THE PATIENT WAS REPORTED TO RECOVER AND HAS BEEN DISCHARGED FROM THE HOSPITAL AT THIS TIME.

    Patient problems: Air Embolism

    MAUDE report 7947729 ↗

FDA clearance and classification

Cleared through 510(k) K050456 on Jun 17, 2005 (decision: Substantially Equivalent). Submitted by Medrad, Inc. as “Medrad Avanta Fluid Management Injection System”. The FDA review took 114 days. FDA record ↗

The FDA regulates product code DXT (injector and syringe, angiographic) under 21 CFR 870.1650 as a Class II device. 21 CFR 870.1650 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050456. One submission often covers many models, so check the product and lot details in each record.

  • Z-2296-2026 · Apr 24, 2026 · Open, Classified
    Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected materi
  • Z-0870-2013 · Nov 5, 2012 · Terminated
    MEDRAD Hand Controller Sheath, Catalog number AVA 500 HCS/3010903, for cardiac catheterization procedures. The Sterile Sheath is specifically intended for use in the x-ray angiogr
  • Z-2866-2011 · Jun 2, 2011 · Terminated
    MEDRAD Avanta Fluid Management Injection System, Catalog numbers: AVA 500 TABL, AVA 500 PEDL Product Usage: Administration of intravascular radio-opaque contrast media and commo
  • Z-1199-2010 · Jan 6, 2010 · Terminated
    Avanta Fluid Management Injection System, Avanta Multi-Patient Disposable Set (MPAT) High Pressure Check Valve, Catalog number AVA 500 MPAT, Material number 3018231.
  • Z-0082-06 · Sep 20, 2005 · Terminated
    Medrad Avanta Single Patient Disposable Sets. Catalog number AVA 500 SPAT
  • Z-0083-06 · Sep 20, 2005 · Terminated
    Medrad Avanta Multi-patient Disposable Sets. Catalog number AVA 500 MPAT

Common questions about the MEDRAD® Avanta Fluid Management Injection System

Has the MEDRAD® Avanta Fluid Management Injection System been recalled?

No FDA recall record in our data names this exact device identifier (DI 00616258008578), based on the FDA recall file dated Nov 27, 2026. However, 6 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the MEDRAD® Avanta Fluid Management Injection System?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 9, 2018 and Nov 18, 2025. In those reports, the most common patient problems are “Air Embolism” (5) and “Cardiac Arrest” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MEDRAD® Avanta Fluid Management Injection System?

1 report is classified as a death, 4 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MEDRAD® Avanta Fluid Management Injection System FDA approved?

It was cleared through 510(k) K050456 on Jun 17, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the MEDRAD® Avanta Fluid Management Injection System, and is it still sold?

The FDA's device database lists Bayer Medical Care Inc. as the company responsible for it (brand MEDRAD® Avanta Fluid Management Injection System, model Avanta). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Injector And Syringe, Angiographic devices

All Injector And Syringe, Angiographic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page