MEDRAD® Stellant FLEX Syringe Kit FLEXD-150-SCSRecalls & reported problems
Bayer Medical Care Inc. · STLNT,KIT,SYR,CT,D,SCS,FLS3,150ML,WLD (87078817) · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00616258022888), received between Oct 30, 2020 and Nov 4, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (2 reports).
1 FDA recall record name this device's identifier.
1 other recall cover devices cleared under the same FDA submission (K173773); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Oct 5, 2020Terminated
Specific lots of MEDRAD Stellant FLEX 150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.
Z-0483-2021 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Manufacturing, Packaging or Shipping Problem2
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 4, 2021Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
THE CUSTOMER REPORTED THE FOLLOWING: AFTER OPENING THE STELLANT FLEX CT DISPOSABLE KIT AND UPON INSPECTION, THEY SAW A FOREIGN BODY WITHIN THE LOW PRESSURE CONNECTOR TUBING (LPCT). THE CUSTOMER ELECTED NOT TO USE THE TUBING. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12752547 ↗MalfunctionOct 26, 2021Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
THE CUSTOMER REPORTED THE FOLLOWING: A FOREIGN MATERIAL WAS OBSERVED IN A STELLANT FLEX SYRINGE. NO INJURY OR ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12697905 ↗MalfunctionOct 30, 2020Leak/Splash
MEDRAD STELLANT FLEX SYRINGES ARE CONSTANTLY LEAKING CONTRAST WHEN IN USE. CONTACTED (B)(4) AND STATED LOT 8586385 HAS KNOWN ISSUES, AND ALL SUPPLY SHOULD BE PULLED FROM BEING USED. FDA SAFETY REPORT ID# (B)(4).
Patient problems: Insufficient Information
MAUDE report 10770196 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173773. One submission often covers many models, so check the product and lot details in each record.
- Z-0745-2019 · Dec 14, 2018 · TerminatedMAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Inje
Other Injector And Syringe, Angiographic devices
- Monarch® 00884450298022Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls37 reports
- Monarch® 00884450298060Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- Blue Diamond™ 00884450498743Merit Medical Systems, Inc. · Injector and syringe, angiographic2 recalls0 reports
- Monarch® 00884450298053Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- NAMIC 80085007Medline Industries, Inc. · Injector and syringe, angiographic1 recall36 reports
- NA 622510B. Braun Medical Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC85 reports
- NAMIC 80087007Medline Industries, Inc. · Injector and syringe, angiographic1 recall20 reports
- Blue Diamond™ 00884450381168Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC1 recall6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.