Zero Gravity ZGM-6-5HRecalls & reported problems
Tidi Products, LLC · Radiation Protection System Floor Unit · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00618125176250), received between Feb 3, 2023 and Sep 24, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement3
- Corroded1
- Device Handling Problem1
- Fracture1
- Inadequate or Insufficient Training1
- Material Puncture/Hole1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Bone Fracture(s)1
- Bruise/Contusion1
- Cramp(s) /Muscle Spasm(s)1
- Insufficient Information1
- Muscle/Tendon Damage1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Material Puncture/Hole
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50891981 ↗MalfunctionMar 18, 2026Unintended Movement
THE SCREWS WERE SHEARED OFF WHEN MOVING THE UNIT BY UNKNOWN AND THE BOOM ARM ROTATED INJURING 2 PEOPLE (UNKNOWN WHO AND WHAT INJURY) BUT DID NOT MENTION THE UNIT TIPPING OVER.
Patient problems: Insufficient Information
MAUDE report 24628623 ↗MalfunctionFeb 11, 2026Fracture
1/15/26-CUSTOMER ASKED FOR QUOTE TO REPLACE THE ZERO GRAVITY BODY SHIELD AS IT DID NOT PASS THE X-RAY CHECK AS IT WAS DISCOVERED THAT THERE IS A LARGE SPLIT IN THE LEAD MODEL: ZGM-6-5H SERIAL#: (B)(6) MANUFACTURED 11-2016
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24318649 ↗MalfunctionDec 30, 2025Mechanical Problem; Mechanical Problem
CUSTOMER ASKED FOR QUOTE FOR A REPLACEMENT ZG BALANCER, SHE STATED THERE IS SOME RESISTANCE WHEN YOU ARE TRYING TO RETRACT THE LEAD APRON. NO DAMAGE TO THE SHEATHING AROUND THE CABLE NOTICED BY THE CUSTOMER. CABLE WAS UPGRADED ON 2/12/21 MODEL: ZGM-6-5H, SERIAL#: (B)(6), MANUFACTURED 3-2014.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23929577 ↗InjuryJul 1, 2025Inadequate or Insufficient Training; Device Handling Problem; Inadequate or Insufficient Training; Device Handling Problem
2 STAFF MEMBERS GOT INJURED TRYING TO MOVE A ZG FLOOR SYSTEM. THE 2 EMPLOYEES HAVE NO EXPERIENCE WITH ZERO GRAVITY AND ATTEMPTED TO MOVE THE SYSTEM WITHOUT THE APPROPRIATE TRAINING ACCORDING TO THE CONTACT AT THE HOSPITAL. THE SYSTEM HASN'T BEEN USED FOR QUITE SOME TIME SO THERE IS A LACK OF KNOWLEDGE ON HOW TO MOVE AND USE THE DEVICE. THERE WERE NO REPORTS OF THE SYSTEM BEING DEFECTIVE. EMPLOYEE 1 SUSTAINED A SERIOUS TOE INJURY DUE TO THE SYSTEM ROLLING/PUSHED OVER HER TOE. THE 2ND EMPLOYEE SUSTAINED A CALF STRAIN TRYING TO PUSH THE ZERO GRAVITY. BUTLER WAS INFORMED THAT TIDI WOULD SEND SOMEONE THERE TO DO AN IN-SERVICE AND TRAIN THEM HOW TO MOVE AND OPERTATE THE SYSTEM.
Patient problems: Bruise/Contusion; Cramp(s) /Muscle Spasm(s)
MAUDE report 22378223 ↗InjuryApr 19, 2024Unintended Movement; Unintended Movement
ZERO GRAVITY MOBILE SYSTEM TIPPED OVER. THE EXACT DATE IS CURRENTLY UNKNOWN. INFORMATION WAS PROVIDED BY THE DISTRIBUTOR, BIOTRONIK. THIS IS THE FIRST INCIDENT; THE SYSTEM TIPPED OVER INJURING / BREAKING A LADY'S FOOT WHEN MOVING THE UNIT. THE SECOND INCIDENT OCCURRED WHEN THE UNIT BEGAN TO TIP OVER, HOWEVER IT WAS CAUGHT BY THE PERSON MOVING THE SYSTEM AND DID NOT FULLY FALL OVER. NO INJURIES WERE REPORTED FOR THE SECOND INCIDENT.
Patient problems: Bone Fracture(s)
MAUDE report 19149317 ↗MalfunctionJun 9, 2023Corroded; Corroded
5/16/23- THE SWIVEL ASSEMBLY IS RUSTED. MODEL: ZGM-6-5H SERIAL#: (B)(6) MANUFACTURED 2017.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17100955 ↗InjuryFeb 3, 2023Unintended Movement; Unintended Movement
INFORMED VIA PHONE CALL BY THE BIOTRONIK DISTRIBUTOR THE FOLLOWING INFORMATION: ORIGINAL PRODUCT FAILURE WAS NOT REPORTED TO TIDI OR BIOTRONIK - POSSIBLE BROKEN SCREW IN STOP BLOCK. REPAIRED PERFORMED BY CUSTOMER BIOMED. WHILE MOVING THE UNIT BEFORE OR AFTER THE REPAIR, IT FELL OVER. THE PERSON DOING THE MOVING OF SYSTEM HAS REPORTED A SORE BACK.
Patient problems: Muscle/Tendon Damage
MAUDE report 16296618 ↗
FDA clearance and classification
The FDA regulates product code EAJ (apron, leaded) under 21 CFR 892.6500 as a Class I device. 21 CFR 892.6500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Zero Gravity ZGM-6-5H
Has the Zero Gravity ZGM-6-5H been recalled?
No FDA recall record in our data names this exact device identifier (DI 00618125176250), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Zero Gravity ZGM-6-5H?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Feb 3, 2023 and Sep 24, 2026. In those reports, the most common specific device problems are “Unintended Movement” (3), “Corroded” (1) and “Device Handling Problem” (1); the most common patient problems are “Bone Fracture(s)” (1), “Bruise/Contusion” (1) and “Cramp(s) /Muscle Spasm(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Zero Gravity ZGM-6-5H?
None of the reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Zero Gravity ZGM-6-5H FDA approved?
Devices of this type (product code EAJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Tidi Products, LLC.
Is the Zero Gravity ZGM-6-5H MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Zero Gravity ZGM-6-5H, and is it still sold?
The FDA's device database lists Tidi Products, LLC as the company responsible for it (brand Zero Gravity, model ZGM-6-5H). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Apron, Leaded devices
- DENTAL DENTAL-RL100-PBurlington Medical, LLC · Apron, Leaded1 recall0 reports
- W113 S8W113*FBurlington Medical, LLC · Apron, Leaded1 recall0 reports
- Zero Gravity ZGCM-HSATidi Products, LLC · Apron, Leaded3 reports
- Zero Gravity ZGHSATidi Products, LLC · Apron, Leaded1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page