AIRCURVE 28224Recalls & reported problems

Resmed Pty Ltd · AIRCURVE 10 ST-A IVAPS · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00619498282241), received between Sep 10, 2025 and Sep 10, 2025 — classified as 2 deaths.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure2

Patient problems

  • Loss of consciousness2
  • Respiratory Failure2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathSep 10, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED TO RESMED THAT A PATIENT WITH A HISTORY OF HYPOVENTILATION SYNDROME AND OBSTRUCTIVE SLEEP APNEA (OSA) WITH MORBID OBESITY, USING AN AIRCURVE 10 ST-A DEVICE EXPIRED ON (B)(6) 2024 ALLEGEDLY DUE TO A DEVICE MALFUNCTION. IT WAS REPORTED THAT ON THE DAY OF INCIDENT; MULTIPLE POWER OUTAGES OCCURRED AT THE PATIENT'S RESIDENCE. IT WAS ALLEGED THAT DURING THESE POWER OUTAGES, THE DEVICE [?]CYCLED' MULTIPLE TIMES AND MAY HAVE MALFUNCTIONED. SHORTLY THEREAFTER, THE PATIENT DEVELOPED SIGNIFICANT SHORTNESS OF BREATH AND BECAME UNRESPONSIVE. THE PATIENT'S WIFE WAS UNABLE TO IMMEDIATELY CALL EMERGENCY SERVICES DUE TO THE OUTAGE. AFTER APPROXIMATELY 20 MINUTES, POWER WAS RESTORED AND EMERGENCY MEDICAL SERVICES WERE CONTACTED. THE PATIENT PASSED AWAY AT THE HOSPITAL ABOUT 4 HOURS LATER. IT WAS REPORTED THAT THE PATIENT HAD BECOME INCREASINGLY DEPENDENT ON THE DEVICE OVER THE YEARS. THE PATIENT'S WIFE REPORTED THAT THEY WERE NOT MADE AWARE THAT THEY WOULD NEED A BACKUP BATTERY OR 'UNINTERRUPTED POWER SUPPLY' FOR THE DEVICE. IT WAS REPORTED BY THE CORONER'S OFFICE THAT THE PATIENT'S PRIMARY CAUSE OF DEATH WAS HYPOVENTILATION SYNDROME AND OSA WITH RESPIRATORY FAILURE SECONDARY TO A FA

    Patient problems: Loss of consciousness; Respiratory Failure

    MAUDE report 23013726 ↗
  • DeathSep 10, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED TO RESMED THAT A PATIENT WITH A HISTORY OF HYPOVENTILATION SYNDROME AND OBSTRUCTIVE SLEEP APNEA (OSA) WITH MORBID OBESITY, USING AN AIRCURVE 10 ST-A DEVICE EXPIRED ON (B)(6) 2024 ALLEGEDLY DUE TO A DEVICE MALFUNCTION. IT WAS REPORTED THAT ON THE DAY OF INCIDENT; MULTIPLE POWER OUTAGES OCCURRED AT THE PATIENT'S RESIDENCE. IT WAS ALLEGED THAT DURING THESE POWER OUTAGES, THE DEVICE CYCLED' MULTIPLE TIMES AND MAY HAVE MALFUNCTIONED. SHORTLY THEREAFTER, THE PATIENT DEVELOPED SIGNIFICANT SHORTNESS OF BREATH AND BECAME UNRESPONSIVE. THE PATIENT'S WIFE WAS UNABLE TO IMMEDIATELY CALL EMERGENCY SERVICES DUE TO THE OUTAGE. AFTER APPROXIMATELY 20 MINUTES, POWER WAS RESTORED AND EMERGENCY MEDICAL SERVICES WERE CONTACTED. THE PATIENT PASSED AWAY AT THE HOSPITAL ABOUT 4 HOURS LATER. IT WAS REPORTED THAT THE PATIENT HAD BECOME INCREASINGLY DEPENDENT ON THE DEVICE OVER THE YEARS. THE PATIENT'S WIFE REPORTED THAT THEY WERE NOT MADE AWARE THAT THEY WOULD NEED A BACKUP BATTERY OR 'UNINTERRUPTED POWER SUPPLY' FOR THE DEVICE. IT WAS REPORTED BY THE CORONER'S OFFICE THAT THE PATIENT'S PRIMARY CAUSE OF DEATH WAS HYPOVENTILATION SYNDROME AND OSA WITH RESPIRATORY FAILURE SECONDARY TO A FAILE

    Patient problems: Loss of consciousness; Respiratory Failure

    MAUDE report 23013722 ↗

Other Ventilator, Continuous, Non-Life-Supporting devices