Ohio Medical, LLC · FM-T series, 15 LPM, USA, 50PSI, OXYGEN · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00628101503000), received between Feb 21, 2021 and Apr 23, 2025 — classified as 1 death, 1 malfunction.

The most frequently reported specific device problem is “Defective Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Component1
  • Improper Flow or Infusion1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Death1
  • Low Oxygen Saturation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2025Improper Flow or Infusion; Improper Flow or Infusion

    ON MARCH 24, 2025, OHIO MEDICAL RECEIVED A MEDSUN FDA-3500A FORM SUBMITTED BY THE USER FACILITY REGARDING AN INCIDENT THAT OCCURRED ON (B)(6) 2025. THE REPORT STATED, "PATIENT WAS ON BIPAP [BILEVEL POSITIVE AIRWAY PRESSURE]; RN [REGISTERED NURSE] CALLED CHARGE RN TO REPORT PATIENTS O2 SATURATION WAS DROPPING FROM 95% TO 88%. UPON ARRIVAL BIBPAP MACHINE SAID "NO O2 FLOW," CHARGE RN CHECKED THE WALL O2 OUTLET AND DISCOVERED THE O2 FLOW METER IN ROOM WAS NOT WORKING AND SWITCHED O2 FLOW METER ON OTHER SIDE OF WALL." THE AFFECTED DEVICE WAS ORIGINALLY SHIPPED ON JUNE 29, 2020. ACCORDING TO THE REPORT, THE PRODUCT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY, AND NO MEDICAL INTERVENTION WAS REQUIRED TO PREVENT PERMANENT IMPAIRMENT. OHIO MEDICAL REVIEWED OUR COMPLAINT DATABASE AND DETERMINED THAT THIS FAILURE HAD NOT BEEN REPORTED TO US. WE THEN INITIATED (B)(4) TO BRING SUSPECT DEVICE, OHIO MEDICAL AMVEX FLOWMETER (SERIAL #(B)(6)) BACK IN HOUSE FOR EVALUATION. MULTIPLE FOLLOW-UP ATTEMPTS WERE MADE BY THE REGISTERED NURSE PATIENT SAFETY COORDINATOR, (B)(6), TO HAVE THE FLOWMETER RETURNED FOR EVALUATION. DESPITE THESE EFFORTS, THE USER FACILITY DID NOT PROVIDE THE DEVICE AND C

    Patient problems: Low Oxygen Saturation

    MAUDE report 21895100 ↗
  • DeathFeb 21, 2021Defective Component; Therapeutic or Diagnostic Output Failure

    OHIO MEDICAL WAS NOTIFIED BY (B)(6) OF THE SUPPLY CHAIN OPERATIONS DEPARTMENT AT (B)(6) MEDICAL CENTER THROUGH OUR NORMAL COMPLAINT COLLECTION PROCESS ON (B)(6) 2021, OF AN ADVERSE EVENT THAT OCCURRED INVOLVING A FLOWMETER WITH POWER TAKE-OFF ADAPTER ASSEMBLY. THE BRIEF DESCRIPTION OF THE INCIDENT PROVIDED WAS "POWER TAKE-OFF WAS NOT FUNCTIONING". OHIO MEDICAL SERVICE ADMINISTRATOR (B)(4) ATTEMPTED TO CONTACT (B)(6) MEDICAL CENTER ON (B)(4) 2021, BY EMAIL FOR FURTHER INFORMATION BECAUSE THE COMPLAINT INDICATED THAT THE PRODUCT CONTRIBUTED TO DEATH, ILLNESS, OR SERIOUS INJURY, AND ALSO TO DETERMINE WHEN THE DEVICE WOULD BE RETURNED FOR EVALUATION. THESE ATTEMPTS WERE UNSUCCESSFUL AS ALL CONTACTS AT (B)(6) MEDICAL CENTER FAILED TO RESPOND. ON (B)(4) 2021, OHIO MEDICAL SR. QUALITY ASSURANCE MANAGER (B)(4) SPOKE WITH THE RISK MANAGER, (B)(6) AT (B)(6) MEDICAL CENTER AND SHE CONFIRMED THAT THE FLOWMETER ASSEMBLY DID NOT CAUSE OR CONTRIBUTE TO THE ADVERSE EVENT. THIS TELEPHONE CONVERSATION WAS SUBSEQUENTLY CONFIRMED ON (B)(4) 2021, VIA EMAIL BY (B)(6) MEDICAL CENTER RISK MANAGER (B)(6). ON FEBRUARY 10, 2021, OHIO MEDICAL RECEIVED USER FACILITY MEDWATCH REPORT (B)(4) VIA (B)(6). MEDWATCH

    Patient problems: Death

    MAUDE report 11356688 ↗