ADVIA Centaur® HBcT 10309508Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · HBc Total (HBcT) Assay Kit 200T · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00630414498096), received between Oct 29, 2019 and Feb 7, 2024 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (6 reports).
2 other recalls cover devices cleared under the same FDA submission (P040004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result6
Patient problems
- Hepatitis3
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2024False Negative Result; False Negative Result
THE CUSTOMER OBTAINED A NON-REACTIVE HBC TOTAL (HBCT) RESULT ON ONE PATIENT SAMPLE TESTED WITH LOT 188 ON AN ADVIA CENTAUR XPT INSTRUMENT, THAT WAS CONSIDERED DISCORDANT RELATIVE TO A REACTIVE RESULT ON A DIFFERENT SAMPLE FROM THE SAME PATIENT USING AN ALTERNATE TESTING METHOD. THE INITIAL ADVIA CENTAUR XPT HBCT NON-REACTIVE RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE PHYSICIAN BELIEVED THE REACTIVE RESULT ON THE ALTERNATE METHOD TO BE THE CORRECT RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18658542 ↗MalfunctionFeb 7, 2024False Negative Result; False Negative Result
THE CUSTOMER OBTAINED A NON-REACTIVE HBC TOTAL (HBCT) RESULT ON ONE PATIENT SAMPLE TESTED WITH LOT 188 ON AN ADVIA CENTAUR XPT INSTRUMENT, THAT WAS CONSIDERED DISCORDANT RELATIVE TO A REACTIVE RESULT ON A DIFFERENT SAMPLE FROM THE SAME PATIENT USING AN ALTERNATE TESTING METHOD. THE INITIAL ADVIA CENTAUR XPT HBCT NON-REACTIVE RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE PHYSICIAN BELIEVED THE REACTIVE RESULT ON THE ALTERNATE METHOD TO BE THE CORRECT RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18658532 ↗MalfunctionNov 16, 2023False Negative Result; False Negative Result
THE CUSTOMER REPORTED OBSERVATION OF ANTI-HEPATITIS B CORE TOTAL (HBCT) NON-REACTIVE (NEGATIVE) RESULTS ON THREE DIFFERENT PATIENT(S) SAMPLES ON AN ADVIA CENTAUR XP INSTRUMENT. THE SAMPLES WERE REPEATED ON THE SAME ADVIA CENTAUR XP INSTRUMENT, AND THE RESULTS WERE NON-REACTIVE (NEGATIVE). THE ORIGINAL HBCT RESULTS WERE REPORTED TO THE PHYSICIAN(S). UPON REPEAT TESTING THE SAMPLES WITH ANTI-HEPATITIS B CORE TOTAL 2 (HBCT2) ASSAY ON THE SAME ADVIA CENTAUR XP INSTRUMENT, THE RESULTS OBTAINED WERE REACTIVE (POSITIVE). THESE RESULTS WERE GENERATED AS PART OF A METHOD COMPARISON BEFORE IMPLEMENTATION OF THE ASSAY AND THE RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). IT IS UNKNOWN WHICH RESULTS ARE CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE NON-REACTIVE (NEGATIVE) HBCT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18151919 ↗InjuryOct 29, 2019False Negative Result; False Negative Result
A CUSTOMER ALLEGED THAT A (B)(6) ADVIA CENTAUR XP HEPATITIS B CORE TOTAL (ANTI-HBCT) RESULT WAS OBTAINED ON (B)(6) 2016 FOR A PATIENT (NON-CADAVERIC SAMPLE) WHO WAS A LIVER TRANSPLANT DONOR. THE COMPLAINT INDICATES THAT A LIVER FROM THIS DONOR WAS TRANSPLANTED TO A RECIPIENT WHO SUBSEQUENTLY DEVELOPED ACUTE (B)(6) INFECTION. THE DONOR SAMPLE WAS THEN RETESTED ON AN ALTERNATE METHOD ON (B)(6) 2019 AND A SECOND ALTERNATE METHOD ON (B)(6) 2019 AND WEAK (B)(6) RESULTS WERE OBTAINED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION FROM THE CUSTOMER, INCLUDING DONOR AND RECIPIENT CLINICAL HISTORY. THE INFORMATION ATTRIBUTED TO THIS EVENT IS DERIVED FROM INFORMATION SUBMITTED TO THE SIEMENS COMPLAINT HANDLING SYSTEM AND HAS NOT BEEN VERIFIED.
Patient problems: Hepatitis
MAUDE report 9253138 ↗InjuryOct 29, 2019False Negative Result; False Negative Result
A CUSTOMER ALLEGED THAT A (B)(6) ADVIA CENTAUR XP HEPATITIS B CORE TOTAL (ANTI-HBCT) RESULT WAS OBTAINED ON (B)(6) FOR A PATIENT (NON-CADAVERIC SAMPLE) WHO WAS A LIVER TRANSPLANT DONOR. THE COMPLAINT INDICATES THAT A LIVER FROM THIS DONOR WAS TRANSPLANTED TO A RECIPIENT WHO SUBSEQUENTLY DEVELOPED ACUTE HBV INFECTION. THE DONOR SAMPLE WAS THEN RETESTED ON AN ALTERNATE METHOD ON (B)(6) 2019 AND A SECOND ALTERNATE METHOD ON (B)(6) 2019 AND WEAK POSITIVE RESULTS WERE OBTAINED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION FROM THE CUSTOMER, INCLUDING DONOR AND RECIPIENT CLINICAL HISTORY. THE INFORMATION ATTRIBUTED TO THIS EVENT IS DERIVED FROM INFORMATION SUBMITTED TO THE SIEMENS COMPLAINT HANDLING SYSTEM AND HAS NOT BEEN VERIFIED.
Patient problems: Hepatitis
MAUDE report 9253137 ↗InjuryOct 29, 2019False Negative Result; False Negative Result
SIEMENS WAS INFORMED THAT A PATIENT WHO HAD PREVIOUSLY BEEN (B)(6) FOR (B)(6) DEVELOPED (B)(6) DURING TRANSPLANT RELATED IMMUNOSUPPRESSION AFTER AN AUTOLOGOUS BONE MARROW TRANSPLANT (PATIENT IS DONOR AND RECIPIENT). SIEMENS HAS REQUESTED ADDITIONAL INFORMATION FROM THE CUSTOMER, INCLUDING DETAILED CLINICAL HISTORY OF THE PATIENT. THE INFORMATION ATTRIBUTED TO THIS EVENT IS DERIVED FROM INFORMATION SUBMITTED TO THE SIEMENS COMPLAINT HANDLING SYSTEM AND HAS NOT BEEN VERIFIED.
Patient problems: Hepatitis
MAUDE report 9253132 ↗
FDA clearance and classification
Its record lists premarket approval application P040004, decided on Dec 22, 2004 with FDA decision code APRL (not a standard approval code). Check the FDA record for its current status. FDA record ↗
The FDA regulates product code SEI (qualitative and quantitative hepatitis b virus antibody assays) under 21 CFR 866.3173 as a Class II device. 21 CFR 866.3173 ↗
FDA definition of this device type: In vitro diagnostic devices intended for use in the detection of antibodies to HBV. These devices are intended to aid in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for HBV infection. Anti-HBs assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown.
Recalls under the same FDA submission
These recalls involve devices cleared under P040004. One submission often covers many models, so check the product and lot details in each record.
- Z-0103-2015 · Sep 15, 2014 · TerminatedADVIA Centaur Systems HBc Total (HBcT) Kit. For In Vitro Diagnostic Use. Catalog Number: 07566733 Siemens Material Number: 10309508 The ADVIA Centaur HBc Total assay is an in
- Z-0102-2015 · Sep 15, 2014 · TerminatedSiemens ADVIA Centaur Systems HBc Total (HBcT), Reagent. For In Vitro Diagnostic Use Catalog Number: 07566733 Siemens Material Number: 10309508 The ADVIA Centaur HBc Total as
Common questions about the ADVIA Centaur® HBcT 10309508
Has the ADVIA Centaur® HBcT 10309508 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00630414498096), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ADVIA Centaur® HBcT 10309508?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Oct 29, 2019 and Feb 7, 2024. In those reports, the most common specific device problem is “False Negative Result” (6); the most common patient problem is “Hepatitis” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ADVIA Centaur® HBcT 10309508?
None of the reports are classified as deaths, 3 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ADVIA Centaur® HBcT 10309508, and is it still sold?
The FDA's device database lists Siemens Healthcare Diagnostics Inc. as the company responsible for it (brand ADVIA Centaur® HBcT, model 10309508). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Qualitative and Quantitative Hepatitis B Virus Antibody assays devices
- ADVIA Centaur® aHBs2 10286268Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays68 reports
- Atellica™ IM aHBs2 10995453Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays58 reports
- ADVIA Centaur aHBe2 50T Kit 10720831Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays1 recall0 reports
- Atellica IM HBcT2 11200739Siemens Healthcare Diagnostics Products Ltd · Qualitative and Quantitative Hepatitis B Virus Antibody assays17 reports
- ADVIA Centaur® aHBs2 10698706Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays13 reports
- Atellica™ IM HBcT 10995597Siemens Healthcare Diagnostics Inc. · Qualitative and Quantitative Hepatitis B Virus Antibody assays8 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page