Atellica IM TSH3-UL 10995703Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · TSH3-UL Atellica IM - RGT 100T · Class 2
All models of Atellica IM TSH3-UL (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00630414600161), received between Jun 3, 2021 and Aug 18, 2023 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (14 reports).
1 other recall cover devices cleared under the same FDA submission (K150403); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Low Test Results14
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2023Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT THAT WAS CONSIDERED DISCORDANT COMPARED TO REPEAT TESTING. THE RESULT WAS REPORTED TO THE PHYSICIAN(S). IT IS UNKNOWN WHETHER THE PHYSICIAN QUESTIONED THE RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17571411 ↗MalfunctionSep 13, 2022Low Test Results; Low Test Results
A DISCORDANT, DEPRESSED THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA IM 1600 ANALYZER COMPARED TO THE RESULT FROM AN ALTERNATE METHOD. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED TSH3-UL PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15410946 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406188 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406183 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406156 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406149 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULTS WHICH WERE DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED TWO INSTANCES WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULTS WERE CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WERE ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406139 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406131 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406121 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULTS WHICH WERE DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED TWO INSTANCES WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULTS WERE CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WERE ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406115 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406107 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULTS WHICH WERE DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED TWO INSTANCES WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULTS WERE CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WERE ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406078 ↗MalfunctionSep 12, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS OBSERVATION OF A FALSELY-DEPRESSED ATELLICA IM THYROID STIMULATING HORMONE 3-ULTRA (TSH3-UL) RESULT WHICH WAS DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. THE CUSTOMER HAS OBSERVED A CASE WHERE THE ATELLICA IM TSH3-UL RESULT IS BELOW THE ASSAY'S MEASURING INTERVAL (<0.008 MIU/L), BUT TESTING USING AN ALTERNATE METHOD PRODUCED A HIGHER RESULT. THE ALTERNATE-METHOD RESULT WAS CONSIDERED CONSISTENT WITH THE PATIENT'S CLINICAL PRESENTATION, AND WAS ACCEPTED AS CORRECT. THERE ARE NO ALLEGATIONS OF PATIENT HARM, AND NO REPORTS OF MODIFICATION OF TREATMENT IN ASSOCIATION WITH THE OBSERVED RESULT DISCORDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406076 ↗MalfunctionJun 3, 2021Low Test Results; Low Test Results
DISCORDANT, FALSELY DEPRESSED THYROID STIMULATING HORMONE 3-ULTRA (TSH3 UL) RESULTS WERE OBTAINED ON PATIENT SAMPLES ON AN ATELLICA IM 1300 (S/N: (B)(4)) INSTRUMENT. THE FALSELY DEPRESSED RESULTS WERE REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON THE SAME INSTRUMENT, GENERATING HIGHER RESULTS WHICH WERE REPORTED TO AND ACCEPTED BY THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED TSH3 UL RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11929516 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K150403. One submission often covers many models, so check the product and lot details in each record.
- Z-2054-2018 · Dec 18, 2017 · Open, ClassifiedADVIA Centaur TSH3-Ultra (Thyroid Stimulating Hormone 3 Ultra), Catalog Numbers: 06491072 (100 tests), 06491080 (500 tests), 04862625 (2500 tests)
Other RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE devices
- VITROS 1912997Ortho Clinical Diagnostics · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE201 reports
- Alinity 07P4830Abbott Ireland · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE2 recalls91 reports
- Elecsys TSH 08443432190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE116 reports
- ARCHITECT 7K62-30Abbott Ireland · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE58 reports
- Elecsys TSH Assay 11731459122Roche Diagnostics GmbH · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE55 reports
- Access TSH (3rd IS) C28643Beckman Coulter, Inc. · Radioimmunoassay, Thyroid-Stimulating Hormone1 recall0 reports
- Alinity 07P4820Abbott Ireland · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE44 reports
- ARCHITECT 7K62-35Abbott Ireland · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page