Vista COV2T Assay 11417414Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Vista COV2T Assay · Class N
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00630414605500), received between Mar 26, 2021 and May 14, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- False Negative Result1
- False Positive Result1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2021False Positive Result; False Positive Result
THE CUSTOMER CONSIDERED A POSITIVE DIMENSION VISTA SARS-COV-2 TOTAL ANTIBODY (COV2T) RESULT ON A PATIENT SAMPLE PROCESSED ON A DIMENSION VISTA 1500 SYSTEM TO BE DISCORDANT. THE PATIENT RESULT WAS REPORTED TO THE PHYSICIAN. THE SAME PATIENT SAMPLE WAS PROCESSED ON A NON-SIEMENS INSTRUMENT SYSTEM USING AN IMMUNGLOBULIN G (IGG) METHODOLOGY. A NEGATIVE RESULT WAS OBTAINED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DIMENSION VISTA COV2T PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11824527 ↗MalfunctionMar 26, 2021False Negative Result; False Negative Result
THE CUSTOMER OBTAINED A DISCORDANT NONREACTIVE (NEGATIVE) DIMENSION VISTA SARS-COV-2 TOTAL ANTIBODY (COV2T) RESULT ON A PATIENT SAMPLE ON A DIMENSION VISTA 500 SYSTEM. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED FOR CONFIRMATION OF THE COV2T RESULT ON A LATER DATE AFTER THE INITIAL RESULT WAS QUESTIONED BY THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED ON THE SAME DIMENSION VISTA SYSTEM AND ON A SECOND DIMENSION VISTA SYSTEM. THE RESULT WAS NEGATIVE ON BOTH SYSTEMS. THE SAME SAMPLE WAS THEN PROCESSED WITH AN ALTERNATE SIEMENS METHODOLOGY ON A SIEMENS ATELLICA SYSTEM AND A REACTIVE (POSITIVE) RESULT WAS OBTAINED. A CORRECTED REPORT WAS ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DIMENSION VISTA COV2T RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11573997 ↗
Other Reagent, coronavirus serological devices
- Access SARS-CoV-2 IgG C58961Beckman Coulter, Inc. · Reagent, Coronavirus Serological57 reports
- VITROS 6199960Ortho Clinical Diagnostics · Reagent, coronavirus serological1 recall0 reports
- VITROS 6199976Ortho-Clinical Diagnostics, Inc. · Reagent, coronavirus serological1 recall0 reports
- Access SARS-CoV-2 IgG QC C58964Beckman Coulter, Inc. · Reagent, Coronavirus Serological1 recall0 reports
- Access SARS-CoV-2 IgG II C69057Beckman Coulter, Inc. · Reagent, Coronavirus Serological33 reports
- ARCHITECT 06R8620Abbott Laboratories · Reagent, coronavirus serological16 reports
- Access SARS-CoV-2 IgM C58957Beckman Coulter, Inc. · Reagent, Coronavirus Serological13 reports
- ARCHITECT 06R8630Abbott Laboratories · Reagent, coronavirus serological3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page