IMMULITE® 2000 Systems CAL Calcitonin 10381446Recalls & reported problems
Siemens Healthcare Diagnostics Products Ltd · Calcitonin - IMMULITE - RGT - 200 Tests · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00630414961156), received between Jan 8, 2018 and Nov 26, 2024 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K023304); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results5
- Incorrect, Inadequate or Imprecise Result or Readings2
- Adverse Event Without Identified Device or Use Problem1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient3
- Complaint, Ill-Defined1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THE OBSERVATION OF CALIBRATION AND REPRODUCIBILITY ISSUES ON PATIENT RESULTS WITH CALCITONIN (CAL), LOT 340 ON AN IMMULITE® 2000 XPI IMMUNOASSAY SYSTEM, SERIAL NUMBER: (B)(6). THE PATIENT RESULTS PROVIDED TO SIEMENS WERE NOT REPORTED TO THE PHYSICIAN AS IT IS UNKNOWN WHICH RESULT IS CORRECT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20779070 ↗MalfunctionMay 17, 2024Non Reproducible Results; Non Reproducible Results
THE CUSTOMER REPORTED IMPRECISE CALCITONIN RESULTS WERE OBTAINED ON TWO PATIENT SAMPLES ON AN IMMULITE XPI INSTRUMENT. THE SAMPLES WERE REPEATED ON THE SAME ANALYZER AND ON AN ALTERNATE IMMULITE 2000 XPI AND THE TRUE RESULT OF BOTH SAMPLES WAS DETERMINED TO BE 10.5 PG/ML. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE IMPRECISE CALCITONIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19341873 ↗MalfunctionSep 9, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER OBSERVED DISCORDANT PATIENT SAMPLE RESULTS WHEN USING IMMULITE 2000 XPI CALCITONIN (CAL) LOT 325, IN COMPARISON TO REPEAT RESULTS. RESULTS FOR THIS SAMPLE WERE NOT REPORTED TO THE PHYSICIAN. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT IMMULITE 2000 XPI CALCITONIN (CAL) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15393421 ↗MalfunctionApr 5, 2022Low Test Results; Low Test Results
THE CUSTOMER OBSERVED DEPRESSED IMMULITE 2000 XPI CALCITONIN RESULTS FROM TWO SAMPLES FROM THE SAME PATIENT UPON COMPARISON WITH REPEAT RESULTS FROM AN ALTERNATE PLATFORM. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THERE WERE NO REPORTS THAT TREATMENT WAS ALTERED OR POSTPONED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DEPRESSED IMMULITE 2000 XPI CALCITONIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14014476 ↗MalfunctionApr 5, 2022Low Test Results; Low Test Results
THE CUSTOMER OBSERVED DEPRESSED IMMULITE 2000 XPI CALCITONIN RESULTS FROM TWO SAMPLES FROM THE SAME PATIENT UPON COMPARISON WITH REPEAT RESULTS FROM AN ALTERNATE PLATFORM. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THERE WERE NO REPORTS THAT TREATMENT WAS ALTERED OR POSTPONED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DEPRESSED IMMULITE 2000 XPI CALCITONIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14014475 ↗MalfunctionNov 15, 2019Low Test Results; Low Test Results
TWO DISCORDANT, FALSELY DEPRESSED CALCITONIN (CAL) RESULTS WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS WERE REPORTED AND QUESTIONED BY THE PHYSICIAN(S). THE SAMPLES WERE REPEATED USING ANOTHER IMMULITE 2000 XPI. THE REPEAT RESULTS WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY DEPRESSED CAL RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9331230 ↗MalfunctionJun 21, 2019Low Test Results; Low Test Results
TWO DISCORDANT, FALSELY DEPRESSED CALCITONIN (CAL) RESULTS WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S). ONE SAMPLE WAS REPEATED USING THE SAME IMMULITE 2000 XPI AND WAS DISCORDANT. ALL FOUR SAMPLES WERE REPEATED USING SIEMENS ADVIA CENTAUR AND THE NEW RESULTS WERE CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED CAL RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8724316 ↗MalfunctionJun 21, 2019Low Test Results; Low Test Results
ONE DISCORDANT, FALSELY DEPRESSED CALCITONIN (CAL) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING SIEMENS ADVIA CENTAUR AND THE NEW RESULT WAS CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED CAL RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8724311 ↗InjuryJan 8, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT DISCORDANT, FALSELY ELEVATED CALCITONIN RESULTS WERE OBTAINED ON ONE PATIENT SAMPLE UPON INITIAL AND REPEAT TESTING ON AN IMMULITE 2000 XPI INSTRUMENT, WHILE USING REAGENT LOT 267. ACCORDING TO THE CUSTOMER: THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S); THE PHYSICIAN(S) PERFORMED AN UNSPECIFIED SURGERY ON THE PATIENT; AND THE PHYSICIAN(S) QUESTIONED THE RESULT OBTAINED ON THE IMMULITE 2000 XPI INSTRUMENT AS THE ELEVATED CALCITONIN RESULT DID NOT MATCH THE CLINICAL PICTURE OF THE PATIENT OBSERVED AFTER SURGERY. THE CUSTOMER REPORTED THAT: IT TESTED THE SAMPLE TWICE ON AN ALTERNATE PLATFORM (RADIOIMMUNOASSAY (RIA) CALCITONIN METHOD); THE RESULTS FROM THE ALTERNATE PLATFORM WERE LOWER THAN THE RESULTS OBTAINED ON THE IMMULITE 2000 XPI INSTRUMENT; FOR TROUBLESHOOTING PURPOSES, IT PERFORMED TESTING WITH HETEROPHILIC BLOCKING TUBES (HBT); THESE TESTS ALSO HAD RESULTS THAT WERE LOWER THAN THE INITIAL AND REPEAT RESULTS OBTAINED ON THE IMMULITE 2000 XPI INSTRUMENT; AND THE CORRECTED RESULT FROM THE ALTERNATE PLATFORM WAS REPORTED TO THE PHYSICIAN(S). OTHER THAN THE ALLEGED UNSPECIFIED SURGERY, THERE HAVE BEEN NO REPORTS OF IMPACT TO THE PATIENT OR ADVERSE
Patient problems: Complaint, Ill-Defined
MAUDE report 7171346 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K023304. One submission often covers many models, so check the product and lot details in each record.
- Z-0212-2015 · Oct 20, 2014 · TerminatedIMMULITE Calcitonin Control Module (IMMULITE, IMMULITE 1000, IMMULITE 2000, IMMULITE 2000 XPi) For in vitro diagnostic use with IMMULITE Systems Analyzers for the quantitative me
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page