IMMULITE®2000 Systems 3gAllergy™ Oak Mix 10386105Recalls & reported problems
Siemens Healthcare Diagnostics Products Ltd · Oak Mix (Red, White, Black) -Specific IgE - IMMULITE - RGT · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00630414965970), received between Oct 21, 2025 and Oct 21, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 6, 2025Class IIOpen, Classified
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Z-0491-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2025Low Test Results; Low Test Results
THE CUSTOMER INDICATED THAT THE IMMULITE 2000 OAK MIX SPECIFIC ALLERGEN LOT 627 WAS SCANNED USING A ZEBRA BARCODE SCANNER AND WAS LOADED ONTO AN IMMULITE 2000 XPI INSTRUMENT. THE CUSTOMER CONFIRMED THAT IMMULITE 2000 OAK MIX SPECIFIC ALLERGEN LOT 627 IS NOT IN USE. THE CUSTOMER DID NOT PROVIDE ANY PATIENT DATA AND DID NOT INDICATE THAT ERRONEOUS PATIENT RESULTS WERE OBTAINED. HOWEVER, IF THE ORDER OF TUBES WITHIN THE ALLERGEN HOLDER WEDGE IS INCORRECT THERE IS A POTENTIAL FOR PATIENT SAMPLE RESULTS TO BE INACCURATE FOR ANY ALLERGEN LOADED IN THE SAME ALLERGEN HOLDER WEDGE AS IMMULITE 2000 OAK MIX SPECIFIC ALLERGEN LOT 627. IT IS UNKNOWN IF OTHER ALLERGEN TUBES WERE LOADED IN THE SAME ALLERGEN HOLDER WEDGE WHICH WERE TESTED/REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23352997 ↗
Other SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL devices
- IMMULITE® 2000 Systems SPE Specific IgE Universal Kit 10708840Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 recall2 reports
- Phadia 1000 12-3800-01Phadia AB · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 recall1 report
- IMMULITE®2000 Systems 3gAllergy™ Food Panel 3 10385891Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 recall0 reports
- IMMULITE®2000 Systems 3gAllergy™ Food Panel 5 10385897Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL7 reports
- IMMULITE®2000 Systems 3gAllergy™ White Ash 10386060Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL4 reports
- IMMULITE®2000 Systems 3gAllergy™ Rat 10385672Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 report
- Hytec Probe Cleaning Solution 288197Hycor Biomedical LLC · System, Test, Radioallergosorbent (Rast) Immunological1 report
- IMMULITE®2000 Systems 3gAllergy™ Timothy Grass 10385924Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page