IMMULITE® 2000 Systems UE3 Unconjugated Estriol 10381192Recalls & reported problems

Siemens Healthcare Diagnostics Products Ltd · Estriol unconjugated - IMMULITE - RGT - 200 Tests · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414997162), received between Sep 7, 2022 and Nov 12, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 12, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED FALSELY DEPRESSED UNCONJUGATED ESTRIOL RESULTS FOR SEVEN PATIENT SAMPLES USING KIT LOT 515 ON AN IMMULITE 2000 XPI INSTRUMENT. THE ERRONEOUS RESULTS WERE REPORTED TO AND QUESTIONED BY THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON AN ALTERNATE IMMULITE 2000 XPI USING KIT LOTS 515 AND 518 AND THE TRUE RESULTS WERE STILL NOT KNOWN. THE SAMPLES WERE THEN TESTED USING A NON-SIEMENS INSTRUMENT AND THE RESULTS WERE HIGHER THAN THE IMMULITE 2000 XPI RESULTS. THE RESULTS OBTAINED USING THE NON-SIEMENS INSTRUMENT WERE REPORTED AS THE CORRECT RESULTS TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED UNCONJUGATED ESTRIOL RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20663242 ↗
  • MalfunctionNov 12, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED FALSELY DEPRESSED UNCONJUGATED ESTRIOL RESULTS FOR SEVEN PATIENT SAMPLES USING KIT LOT 515 ON AN IMMULITE 2000 XPI INSTRUMENT. THE ERRONEOUS RESULTS WERE REPORTED TO AND QUESTIONED BY THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON AN ALTERNATE IMMULITE 2000 XPI USING KIT LOTS 515 AND 518 AND THE TRUE RESULTS WERE STILL NOT KNOWN. THE SAMPLES WERE THEN TESTED USING A NON-SIEMENS INSTRUMENT AND THE RESULTS WERE HIGHER THAN THE IMMULITE 2000 XPI RESULTS. THE RESULTS OBTAINED USING THE NON-SIEMENS INSTRUMENT WERE REPORTED AS THE CORRECT RESULTS TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED UNCONJUGATED ESTRIOL RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20663240 ↗
  • MalfunctionSep 7, 2022Low Test Results; Low Test Results

    THE CUSTOMER RAN A PRENATAL SCREENING SAMPLE ON IMMULITE 2000 XPI UNCONJUGATED ESTRIOL (UE3), WHICH WAS OUT OF ADJUSTMENT. THIS LED TO A FALSELY DEPRESSED UE3 RESULT, INDICATING HIGH-RISK FOR TRISOMY 18. UPON RETEST, THE UE3 RESULT WAS HIGHER (LOW RISK FOR TRISOMY 18). IT IS CURRENTLY UNCLEAR WHETHER THE INITIAL FALSLEY DEPRESSED RESULT WAS REPORTIED TO THE PHYSICIAN. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE UE3 RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15372957 ↗