Resolute Integrity™ RSINT30009UXRecalls & reported problems
Medtronic, Inc. · Stent RSINT30009UX MicroTrac 3.00X9RX · Class 3 · Implant
All models of Resolute Integrity (52)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169037489), received between Apr 5, 2017 and May 8, 2018 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Device Damaged by Another Device1
Patient problems
- No Consequences Or Impact To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 8, 2018Device Damaged by Another Device; Device Dislodged or Dislocated; Device Damaged by Another Device; Device Dislodged or Dislocated
DURING THE PROCEDURE AN ATTEMPT WAS MADE TO USE ONE RESOLUTE INTEGRITY RX CORONARY DRUG ELUTING STENT TO TREAT A SEVERELY CALCIFIED AND MILDLY TORTUOUS LESION LOCATED IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY, EXHIBITING 95% STENOSIS. THERE WERE NO ABNORMALITIES IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. A RIGHT RADIAL APPROACH WAS USED. THE LESION WAS PRE-DILATED. THE DEVICE PASSED THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. IT WAS REPORTED THAT DURING THE DELIVERY OF THE STENT TO THE MID LAD, THE DELIVERY CATHETER BROKE JUST DISTAL TO THE LUER/HUB. THE STENT WAS NOT DEPLOYED, AND THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ISSUE. AN ATTEMPT WAS MADE TO USE ANOTHER RESOLUTE INTEGRITY RX CORONARY DRUG ELUTING STENT TO TREAT THE SAME LESION. NO DAMAGE WAS NOTED TO PACKAGING. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION PRE-DILATE
Patient problems: Patient Problem/Medical Problem
MAUDE report 7493627 ↗InjuryApr 5, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PHYSICIAN INTENDED TO USE A RESOLUTE INTEGRITY DRUG-ELUTING STENT TO TREAT A CALCIFIED LESION IN THE RCA WITH A PREVIOUS DEPLOYED DES. NO DAMAGE NOTED TO DEVICE PACKAGING. DEVICE WAS INSPECTED PRIOR TO USE NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED FOLLOWED BY NEUTRAL PREP. LESION WAS PRE-DILATED. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE PAST A JUST DEPLOYED RESOLUTE, EXCESSIVE FORCE WAS NOT USED. PHYSICIAN WAS UNABLE TO DEPLOY THE STENT AND WHEN THE PHYSICIAN PULLED IT BACK INTO THE 6F GUIDELINER, HE NOTICED UNDER FLUORO, THAT THE STENT SLIPPED OFF THE BALLOON. HE THEN REMOVED THE GUIDELINER AND THE STENT WITHOUT ANY ISSUE TO THE PATIENT. HE SUCCESSFULLY DEPLOYED A RESOLUTE IMMEDIATELY AFTERWARDS. PHYSICIAN INSISTS THAT THERE WAS NO ISSUE WITH THE RESOLUTE INTEGRITY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6459553 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page