N/A 5584002Recalls & reported problems

Medtronic Sofamor Danek, Inc. · DRIVER 5584002 NON-BRKOFF PROVSNL INSERT · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169064133), received between Nov 6, 2024 and Nov 18, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PRODUCT USED FOR SPINAL FUSION THERAPY. LEVELS IMPLANTED L4-5. IT WAS REPORTED THAT A PART OF THE SHAFT IN THE MIDDLE OF THE DRIVER WAS BROKE OFF BEFORE THE SURGERY STARTED. THERE WERE NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23581036 ↗
  • MalfunctionJan 25, 2025Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE FACILITY VIA MANUFACTURER REPRESENTATIVE REGARDING TWO DRIVERS USED FOR SPINAL FUSION WITH TRANSFORAMINAL LUMBAR INTERBODY FUSION. IT WAS REPORTED THAT BOTH DRIVERS BROKE. THERE WERE NO FRAGMENTS OF THE IMPLANTS OR INSTRUMENTS REMAINED IN THE PATIENT'S BODY. PRODUCT WILL BE RETURNED. THERE WERE NO FURTHER COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE DRIVERS BROKE WHILE TRYING TO SNAP OFF SET SCREW DURING PROCEDURE. PRE-OP DIAGNOSIS UNKNOWN. PATIENT WAS INVOLVED IN THE EVENT. PRODUCT CAME IN CONTACT WITH THE PATIENT. BUT THERE WERE NO COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21236335 ↗
  • MalfunctionJan 15, 2025Break; Break

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING INSTRUMENTS USED IN TLIF. IT WAS REPORTED THAT THE INSTRUMENTS WERE DAMAGED. THE ITEMS BECAME DAMAGED AND WORN OUT OVER TIME IN SPD. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21155662 ↗
  • MalfunctionJan 15, 2025Break; Break

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING INSTRUMENTS USED IN TLIF. IT WAS REPORTED THAT THE INSTRUMENTS WERE DAMAGED. THE ITEMS BECAME DAMAGED AND WORN OUT OVER TIME IN SPD. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21155432 ↗
  • MalfunctionNov 6, 2024Break; Naturally Worn; Break; Naturally Worn

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (QUALITY INSPECTOR) REGARDING A PART USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE TIP WAS BROKEN AND INSTRUMENT WAS WORN. THERE WAS NO FURTHER INFORMATION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20619834 ↗