Medtronic Sofamor Danek, Inc. · CONTROLLER 945ECLC ECLIPSE · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Eclipse controller is an electrical stimulator for monitoring evoked responses. Reports mostly describe the controller disconnecting, failing to recognize the stim box, or otherwise not working properly. There are 24 reports naming its identifier, with no deaths and 1 injury. No recalls name this device, and reports fell to 2 in the last 12 months from 9 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00643169071872), received between Aug 3, 2017 and Nov 17, 2025 — classified as 1 injury, 23 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 7 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem9
  • Device Sensing Problem4
  • Failure to Sense3
  • Intermittent Continuity3
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Connection Problem1
  • Defibrillation/Stimulation Problem1
  • Device Displays Incorrect Message1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Insufficient Information8
  • No Consequences Or Impact To Patient4
  • No Known Impact Or Consequence To Patient2
  • Burn(s)1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2025Intermittent Continuity; Intermittent Continuity

    IT WAS REPORTED THAT THE NIM ECLIPSE CONTROLLER GETS DISCONNECTED WHEN PLUGGED IN. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 23569168 ↗
  • MalfunctionNov 17, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT NIM ECLIPSE HAD NON-SPECIFIC DAMAGED ISSUE. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23567053 ↗
  • MalfunctionAug 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT CONTROLLER WAS NOT RECOGNIZING STIM BOX. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 22892157 ↗
  • MalfunctionAug 18, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT ECLC CONTROLLER WAS NOT WORKING PROPERLY. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 22836308 ↗
  • MalfunctionAug 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DEVICE WAS NOT WORKING DAQ A DEFECTIVE. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 22824912 ↗
  • MalfunctionAug 18, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE STIM BOX WAS NOT DETECTED. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 22824718 ↗
  • MalfunctionAug 18, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE CONTROLLER HAD BAD STIM BOARD WITH NOT WORKING PROPERLY. THERE WAS NO KNOWN PATIENT IMPACT RELATED TO THE EVENT.

    Patient problems: Insufficient Information

    MAUDE report 22824714 ↗
  • MalfunctionAug 18, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE DEVICE HAD DAQ BOARD OUT. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 22824662 ↗
  • MalfunctionAug 17, 2025Failure to Sense; Failure to Sense

    IT WAS REPORTED THAT THE NIM ECLIPSE HAD CONNECTION FAILURE. THERE WAS NO KNOWN PATIENT IMPACT IN THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 22824572 ↗
  • MalfunctionMar 6, 2025Device Displays Incorrect Message; Device Sensing Problem; Device Displays Incorrect Message; Device Sensing Problem

    IT WAS REPORTED THAT DAQ B DOES NOT APPEAR TO BE RECOGNIZED. AN ERROR MESSAGE WAS DISPLAYED ABOUT DAQ B WHEN MODALITIES RECORDING FROM THAT BOX ARE TURNED ON. ADDITIONALLY, THE IMPEDANCE CHECK ONLY SHOWED GREY BOXES FOR PROGRAMMED INPUTS. THIS ISSUE WAS CONFIRMED WITH DIFFERENT CABLES, DAQ BOXES, AND CONTROLLERS. THE CONTROLLER WAS SWAPPED OUT TO COMPLETE THE CASE. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21539636 ↗
  • MalfunctionDec 30, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    IT WAS REPORTED THAT ECLC CONTROLLER HAD NO STIM RETURN. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21034559 ↗
  • MalfunctionSep 9, 2024Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED DURING INTRA-OP THAT THE SYSTEM WAS NOT RECOGNIZE THE EEX 901S. CABLES AND BOXES WERE SWAPPED, HOWEVER THE ONLY THING THAT RESOLVED IT WAS TO CHANGE THE BASE UNIT. NO ISSUES AROSE WITH THE SAME BOXES AND CABLES WITH A NEW BASE. ADDITIONALLY, THE TECHNOLOGIST HAS REPORTED ONGOING ISSUE WITH 60HZ OUTSIDE OF NORMAL AND WITHOUT A IDENTIFIABLE SOURCE. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20167320 ↗
  • MalfunctionAug 22, 2024Intermittent Continuity; Intermittent Continuity

    IT WAS REPORTED THAT DAW B INTERMITTENTLY DROPS OUT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20028559 ↗
  • MalfunctionMay 24, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE NIM HAD RECUURENT STIM OUTPUT FAILURE. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16990487 ↗
  • MalfunctionMar 6, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THERE WAS A PROCEDURE PERFORMED IN THE OPERATING ROOM DURING WHICH ONE OF THE TECHS WAS USING A MEDTRONIC E4 MACHINE. THE PATIENT REPORTEDLY HAD A POOR OUTCOME, AND THE HOSPITAL HAS REQUESTED THAT A CERTIFIED MEDTRONIC REPRESENTATIVE TEST THE EQUIPMENT TO ENSURE IT IS IN PROPER WORKING CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16485228 ↗
  • MalfunctionMay 25, 2022Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE DEVICE PREAMP B IS NOT WORKING .NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14495651 ↗
  • MalfunctionApr 29, 2022Device Sensing Problem; Device Sensing Problem

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE CONTROLLER DELIVERS NO MOTOR STIM AND INCONSISTENT SENSORY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14253158 ↗
  • InjuryMar 24, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE HCP REPORTED THAT WHILE MONITORING A LUMBER FUSION CASE, AND HAD PLACED AN ANODE NEEDLE ELECTRODE TO USE WITH PROBE FOR SCREW TESTING. WHEN PATIENT WOKE UP, THERE WAS A THIRD DEGREE BURN AT THE SITE OF THE ANODE NEEDLE. ELECTROCAUTERY WAS ALSO USED BY SURGEON DURING PROCEDURE. AFTER THE INTERNAL TESTING RESULTS CAME BACK AS THE MACHINE TESTING NORMAL, THE USER SUSPECT THAT THE ELECTROCAUTERY MACHINE WAS NOT PROPERLY GROUNDED AND FOUND A PATH THROUGH THE IOM PLACED NEEDLE.

    Patient problems: Burn(s)

    MAUDE report 13884725 ↗
  • MalfunctionMar 1, 2021Device Sensing Problem; Device Sensing Problem

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE CONNECTOR FAILED WHEN CONNECTED TO PATIENT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11391813 ↗
  • MalfunctionDec 28, 2018Failure to Sense; Failure to Sense

    A HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT THE CONTROLLER WAS HAVING ISSUES WITH ACQUIRING MOTOR EVOKED POTENTIALS (MEP). NO RESPONSE WAS ACHIEVED AFTER HAVING A VOLTAGE UP UNTIL 700 VOLTS. THE PATIENT WOULD SOMETIMES REACT TO THE STIMULUS AND SOMETIMES THERE IS NO REACTION. NO BACK-UP DEVICE WAS USED AND THERE WERE NO ISSUES DURING THE PROCEDURE BUT MEP WAS NOT RUN AT THE TIME. THERE WAS NO KNOWN PATIENT IMPACT REPORTED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8204017 ↗
  • MalfunctionDec 14, 2018Use of Device Problem; Connection Problem; Use of Device Problem; Connection Problem

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE PREAMPLIFIER B PORT NEEDED REPAIR AND THAT THE DEVICE WAS NOT WORKING PROPERLY DURING THE PROCEDURE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8164326 ↗
  • MalfunctionNov 20, 2018Failure to Sense; Failure to Sense

    A HEALTH CARE PROVIDER (HCP) REPORTED THAT THE DEVICE WAS NOT PRODUCING A HIGH STIMULUS; WHEN BOTH STIMULATORS WERE PLUGGED IN THEY WERE GETTING AN ELECTRICAL STIM FAILURE; TRIED WITH TWO OTHER STIMULATORS ON THE SAME DEVICE AND HAD THE SAME ISSUE. THERE WAS NO KNOWN PATIENT IMPACT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8090442 ↗
  • MalfunctionAug 2, 2018Intermittent Continuity; Loss of Power; Use of Device Problem; Intermittent Continuity; Use of Device Problem

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE CONTROLLER WAS INTERMITTENTLY NOT WORKING AND SHUTS DOWN IN THE MIDDLE OF CASES. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7741995 ↗
  • MalfunctionAug 3, 2017Sensing Intermittently; Sensing Intermittently

    THE FACILITY REPORTED THAT PRE-OPERATIVE THE DEVICE HAD AN INTERMITTENT STIM OUTPUT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6765279 ↗

Other STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices

All STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices →

Sources and updates