CD HORIZON® Spinal System 1555501100Recalls & reported problems

Medtronic Sofamor Danek, Inc. · ROD 1555501100 5.5 CCM NS CURV 100MM · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169081345), received between Nov 2, 2020 and Nov 2, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

1 other recall cover devices cleared under the same FDA submission (K113174); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Biocompatibility1

Patient problems

  • Foreign Body Reaction1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 2, 2020Biocompatibility; Biocompatibility

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH AN UNKNOWN INDICATION IN NEED OF L1-L5 POSTERIOR APPROACH PATIENT PRONE PROCEDURE USED IN SPINAL THERAPY. IT WAS REPORTED THAT THE PATIENT HAD A REACTION POTENTIALLY ALLERGIC REACTION TO THE COBALT CHROME (CHROMALOY) CONSTRUCT OF SOLERA 5.5/6.0. THERE WAS METAL DEBRIS IS SOFT TISSUE AND MUSCLE THAT WAS VISIBLE AFTER INCISION. REVISION SURGERY WAS PERFORMED AND REPLACED ALL SCREWS WITH LEGACY TITANIUM AND EXTENDED THE CONSTRUCT. A PREVIOUS METAL ALLERGY WAS NOT DOCUMENTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Foreign Body Reaction

    MAUDE report 10768674 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113174. One submission often covers many models, so check the product and lot details in each record.

  • Z-2603-2018 · Jul 9, 2018 · Terminated
    Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.

Other Pedicle screw spinal system, adolescent idiopathic scoliosis devices

All Pedicle screw spinal system, adolescent idiopathic scoliosis devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page