Medtronic Reusable Instruments 3990004Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SHAFT 3990004 SHORT THREADED · Class 2
All models of Medtronic Reusable Instruments (36)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169091474), received between Oct 7, 2021 and Mar 2, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Deformation1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 2, 2026Break
INFORMATION WAS RECEIVED FROM THE USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL PROCEDURE. IT WAS REPORTED THAT THE SHAFT WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED R EGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24480239 ↗MalfunctionDec 20, 2021Scratched Material; Scratched Material
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PRE-OPERATIVE DIAGNOSIS FOR KYPHOSIS. PROCEDURE OR TECHNIQUE USED WAS PLIF. LEVELS IMPLANTED L4/5. IT WAS REPORTED THAT THE SURGEON CHOOSE A CAGE OF A NO PROBLEM SIZE WITH THE TRIAL, BUT WHEN THEY INSERTED THE CAGE, THEY COULD NOT TURN THE INSERTER. ONCE THE SURGEON PULLED OUT THE INSERTER, CAGE WAS TWISTED. THERE WAS NO FRAGMENT LEFT IN THE PATIENT. THERE WERE NO SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. THE CAGE WAS TWISTED. INSERTER TIPS WAS SLIGHTLY DEFORMED. SCRATCHES WERE FOUND AT THE TIP, THREADS NEAR THE HANDLE AND BOTTOM OF THE SHAFT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13041190 ↗MalfunctionOct 7, 2021Material Deformation; Material Deformation
INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING INTRA-OP FOR A PATIENT UNDERGOING L2-S2AI-PSF, L4/5-PLIF. THE PRE-OPERATIVE DIAGNOSIS WAS MENTIONED AS LUMBAR SPINAL CANAL STENOSIS. IT WAS REPORTED THAT, THE SHORT INSERTER WAS FOUND TO HAVE A CRACK AT THREAD AND THE OTHER INSERTER SHAFTS WERE NOT ATTACHED TO IMPLANT DUE TO THREAD DAMAGE. THE INSERTER TIP WAS DEFORMED. THE INSERTER SHAFT WAS STRIPPED AT THE TIP AND SCRATCHES AND BURRS WERE FOUND AT THREADS ON HANDLE SIDE. THE OTHER INSERTER HAD SCRATCHES AT TIP THREADS AND SCRATCHES AND STRIPPING FOUND AT THREADS. THE WAS A DELAY OF LESS THAN 60 MINUTES REPORTED AS A RESULT OF THIS EVENT. PATIENT MEDICAL HISTORY INCLUDES HT, DM, CEREBRAL INFARCTION, TWO-SIDED TKA, DVT. THERE WAS NO HEALTH DAMAGE IN PATIENT REPORTED AS A RESULT OF THIS EVENT. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12588964 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page