Contour 3D® 690R30Recalls & reported problems
Medtronic, Inc. · RING 690R30 CONTOUR 3D 26L MEX · Class 2 · Implant
Quick summary
This Medtronic Contour 3D 690R30 is a flexible annuloplasty ring shaped to repair the tricuspid valve of the heart. The 51 reports mostly describe the ring being taken out and replaced during or shortly after surgery — because it was damaged, did not fit or seal properly, or left blood leaking backwards through the valve — along with heart rhythm problems that led some patients to need a pacemaker or defibrillator. Reports include 1 death and 49 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 50
- Deaths reported
- 1
- Injuries reported
- 48
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 00643169121690), received between Jun 26, 2018 and Aug 18, 2026 — classified as 1 death, 48 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (28 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information28
- Adverse Event Without Identified Device or Use Problem10
- Patient Device Interaction Problem10
- Intravalvular regurgitation6
- Inadequacy of Device Shape and/or Size4
- Appropriate Device Problem Term/Code Not Available2
- No Apparent Adverse Event2
- Incomplete Coaptation1
Patient problems
- Insufficient Information22
- Tricuspid Valve Insufficiency/ Regurgitation12
- No Information7
- Tricuspid Regurgitation4
- Non specific EKG/ECG Changes3
- Tachycardia3
- Atrial Fibrillation2
- Heart Block2
Recent adverse event reports
The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.
InjuryAug 18, 2026Insufficient Device Problem Information
Medtronic received information that 12 days post implant of this 30mm tricuspid annuloplasty ring, it was explanted and replaced with a 31mm mitral bioprosthetic valve in the same position. The reason for the replacement was not reported. No additional adverse patient effects were reported.
Patient problems: Insufficient Information
MAUDE report 50303730 ↗InjuryApr 22, 2026Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM CONTOUR 3D ANNULOPLASTY HEART RING, IT WAS EXPLANTED AND REPLACED WITH ANOTHER MDT PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED DUE TO THE RING BEING DAMAGED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24954661 ↗InjuryApr 20, 2026Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS POST IMPLANT OF THIS 30MM CONTOUR 3D ANNULOPLASTY HEART RING, THE PATIENT WAS EXPERIENCING BIFASCICULAR BLOCK, WHICH RESULTED IN HOSPITALIZATION. TELEMETRY WAS PERFORMED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24925756 ↗InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT OF THE 30MM TRICUSPID ANNULOPLASTY RING WAS RESIDUAL RE GURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 24349603 ↗InjuryFeb 16, 2026Structural Problem; Incomplete Coaptation
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 28MM ANNULOPLASTY RING. THE REASON FOR THE REPLACEMENT WAS REPORTED AS IMPROPER LEAFLET COAPTATION AND AND A DEFECT IN THE STRUCTURE OF THE RING. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24345241 ↗InjuryJan 22, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT 4 MONTHS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 24143566 ↗InjuryJan 15, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT 11 MONTHS POST IMPLANT OF THIS 30MM TRICUSPID HEART RING, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS LOW EJECTION FRACTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24078109 ↗InjuryDec 30, 2025Intravalvular regurgitation; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE ANNULOPLASTY RING WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL. IT WAS ALSO NOTED THAT THE REPLACEMENT DEVICE WAS A 29MM BIOPROSTHETIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 23926296 ↗InjuryOct 21, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT 3 DAYS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Heart Block
MAUDE report 23347162 ↗InjurySep 9, 2025Appropriate Device Problem Term/Code Not Available; Patient Device Interaction Problem; Appropriate Device Problem Term/Code Not Available; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 5 MONTHS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY AND HISTORY OF SPONTANEOUS SUSTAINED VENTRICULAR TACHYCARDIA. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiomyopathy; Tachycardia
MAUDE report 23006450 ↗MalfunctionSep 4, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS REPORTED THAT AN ECHOCARDIOGRAM DISCOVERED MODERATE TRICUSPID REGRUGITATION. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 22962611 ↗InjuryAug 26, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN ANNULOPLASTY PRODUCT OF THE SAME SIZE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS THE ANNULOPLASTY DID NOT MATCH THE PATIENT'S TRICUSPID ANNULUS MORPHOLOGY. WATER TESTS WERE COMPLETED, AND THE EFFECT WAS POOR WITH MODERATE TO SEVERE TRICUSPID REGURGITATION AND PROLAPSE. THE ANNULOPLASTY BAND WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 22899021 ↗InjuryJun 17, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 1 WEEK POST IMPLANT OF THIS 28MM PROFILE 3D RING AND 30MM CONTOUR 3D RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 22217484 ↗InjuryMar 10, 2025Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS SEVERE TRICUSPID REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 21566552 ↗InjuryFeb 13, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SINUS BRADYCARDIA AND ATRIAL ARRHYTHMIAS (ATRIAL TACHYCARDIA (AT), ATRIAL FIBRILLATION AND ATRIAL FLUTTER). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Atrial Fibrillation; Bradycardia; Non specific EKG/ECG Changes; Tachycardia
MAUDE report 21381234 ↗DeathJan 2, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A POSTOPERATIVE DEATH OCCURRED FOLLOWING THE IMPLANTATION OF A 30MM CONTOUR 3D TRICUSPID ANNULOPLASTY RING. THE CAUSE OF DEATH IS UNKNOWN. THERE WERE NO ALLEGED PRODUCT ISSUES REPORTED. THE DEVICE REMAINS IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 21063029 ↗InjuryDec 6, 2024Intravalvular regurgitation; Intravalvular regurgitation
IT WAS REPORTED THAT 5 MONTHS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, SEVERE TRICUSPID REGURGITATION WAS NOTED. IT WAS REPORTED THAT PERCUTANEOUS TRICUSPID VALVE RECONSTRUCTION USING TRICLIP IMPLANTATION BETWEEN THE SEPTAL AND POSTERIOR TRICUSPID NATIVE VALVE LEAFLET WAS PERFORMED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 20865389 ↗InjuryOct 15, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 2 DAYS POST IMPLANT OF THIS 30 MM TRICUSPID ANNULOPLASTY RING AND 30 MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS 3RD DEGREE ATRIOVENTRICULAR BLOCK/COMPLETE HEART BLOCK AND ATRIAL ARRHYTHMIAS INDICATED TO BE ATRIAL TACHYCARDIA, ATRIAL FIBRILLATION (AFIB) AND ATRIAL FLUTTER. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Atrial Fibrillation; Atrial Flutter; Heart Block; Tachycardia
MAUDE report 20449253 ↗InjurySep 5, 2024Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A VALVE OF A DIFFERENT MANUFACTURER. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SEVERE TRICUSPID REGURGITATION DUE TO THE PATIENTS NATIVE LEAFLET NOT COAPTING PROPERLY. IT WAS STATED THAT THE 30MM TRICUSPID ANNULOPLASTY RING MALFUNCTIONED. NO ADDITIONALADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 20147947 ↗InjuryMar 18, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Appropriate Device Problem Term/Code Not Available
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS REPORTED TO BE "WASTED" AND "IN/OUT". IT WAS REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 18922266 ↗InjuryMar 14, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18907507 ↗InjuryMar 5, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 6 DAYS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS REPORTED THAT A COMPUTERIZED TOMOGRAPHY (CT) SCAN DISCOVERED A SUBDURAL HEMATOMA. IT WAS ALSO REPORTED THAT THE HEMATOMA WAS LOCATED IN THE FRONTOPARIETAL RIGHT AND FRONTAL LEFT AND PARAFACIAL. AN UNKNOWN INTERVENTION OCCURRED HOWEVER THE DEVICE REMAINS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Hematoma
MAUDE report 18838031 ↗InjuryFeb 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED ADDITIONAL INFORMATION WHICH STATED THAT THE OCCLUSION OF THE RIGHT CORONARY ARTERY WAS NOT DUE TO A MEDTRONIC DE VICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Hematoma; Pericardial Effusion
MAUDE report 18677515 ↗InjuryJan 9, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 7 MONTHS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 42MM ANNULOPLASTY BAND OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18475446 ↗InjurySep 25, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 17808114 ↗
Other RING, ANNULOPLASTY devices
- CG FUTURE™ 638BL32Medtronic, Inc. · RING, ANNULOPLASTY1 death71 reports
- PROFILE 3D® 680R28Medtronic, Inc. · RING, ANNULOPLASTY1 death67 reports
- CG FUTURE™ 638BL30Medtronic, Inc. · RING, ANNULOPLASTY1 death65 reports
- CG FUTURE™ 638RL32Medtronic, Inc. · RING, ANNULOPLASTY59 reports
- Tri-Ad™ 900SFC228Medtronic, Inc. · RING, ANNULOPLASTY4 deaths56 reports
- CG FUTURE™ 638BL34Medtronic, Inc. · RING, ANNULOPLASTY51 reports
- CG FUTURE™ 638RL30Medtronic, Inc. · RING, ANNULOPLASTY1 death47 reports
- CG FUTURE™ 638BL28Medtronic, Inc. · RING, ANNULOPLASTY1 death47 reports