Attain Select™ II + SureValve™ 6248V-90S (UDI 00643169143197)Recalls & reported problems
Medtronic, Inc. · Catheter 6248V-90S Attain Select II · Class 2
All models of Attain Select II + SureValve (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169143197), received between Oct 6, 2017 and Nov 19, 2019 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Material Fragmentation3
- Material Integrity Problem2
- Break1
- Detachment Of Device Component1
- Device Or Device Fragments Location Unknown1
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2019Mechanical Problem; Detachment of Device or Device Component; Material Integrity Problem; Mechanical Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE PHYSICIAN WAS ATTEMPTING TO PLACE THE LEFT VENTRICULAR (LV) LEAD WHEN THE CATHETER WAS UNABLE TO BE ADVANCED DUE TO PATIENT ANATOMY. THE PHYSICIAN ATTEMPTED TO SLIT THE INNER CATHETER BUT WAS NOT ABLE TO SLIT ALL THE WAY THROUGH THE WHITE HUB. THE TIP OF THE CATHETER BROKE OFF. NO LV LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9341897 ↗MalfunctionAug 5, 2019Material Fragmentation; Detachment of Device or Device Component; Material Fragmentation; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING IMPLANT THE CATHETER BROKE DURING SLITTING AND REMOVAL. THE CATHETER WAS REMOVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8862248 ↗MalfunctionMar 22, 2019Material Fragmentation; Detachment of Device or Device Component; Material Fragmentation; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE REPLACEMENT PROCEDURE, THE INNER CATHETER BROKEN IN HALF AND THE LEFT VENTRICULAR (LV) LEAD DISLODGED WHEN SLITTING THE CATHETER. THE LEAD AND THE CATHETER WERE NOT USED, AND ANOTHER LV LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8445152 ↗MalfunctionJun 1, 2018Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT WHILE SLITTING THE INNER CATHETER BROKE. THE CATHETER WAS ABLE TO BE REMOVED WITH THE LEAD INTACT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7560317 ↗MalfunctionOct 6, 2017Break; Detachment Of Device Component; Material Fragmentation; Device Or Device Fragments Location Unknown; Break; Detachment Of Device Component; Material Fragmentation; Device Or Device Fragments Location Unknown
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE WHEN THE PHYSICIAN STARTED TO SLIT THE CATHETER, THE HUB AND PART OF THE PROXIMAL CATHETER TOTALLY SEPARATED AND DETACHED FROM THE REMAINING BODY OF THE CATHETER. THE PHYSICIAN WAS REQUIRED TO CUT AWAY THE REMAINING CATHETER WITH IRIS SCISSORS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6920978 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.