CAPSTONE PTC™ Spinal System 3992206Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SPACER 3992206 22MM X 6MM · Class 2 · Implant
All models of CAPSTONE PTC Spinal System (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169187573), received between Feb 7, 2017 and Oct 20, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Adverse Event Without Identified Device or Use Problem1
- Crack1
- Improper or Incorrect Procedure or Method1
- Positioning Problem1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 20, 2024Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED DURING A T9-IL SPINAL FUSION PROCEDURE IN A PATIENT DIAGNOSED WITH DEGENERATIVE SCOLIOSIS. IT WAS REPORTED THAT WHILE L5-S1 WAS QUITE NARROW, CAGE WAS OVER-INSERTED. AS SUCH, WHEN THE POSITION OF THE CAGE WAS SHIFTED TO THE DORSAL SIDE, IT WAS RETURNED WITH A SLIDE HAMMER, BUT IT DID NOT RETURN PROPERLY; THE DORSAL PORTION OF THE CAGE WAS DAMAGED. THE BACK PART OF THE CAGE WHICH WAS MISSING WAS ATTACHED TO THE INSERTER. NO FRAGMENTS ARE REMAINING IN THE PATIENT. THE CAGE BEING OVER-INSERTED AND GETTING DAMAGED WHEN IT WAS PULLED OUT WAS DUE TO ERROR. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20489857 ↗MalfunctionJun 9, 2020Break; Break
PRE-OPERATIVE DIAGNOSIS: KYPHOSIS PROCEDURE PERFORMED: POSTERIOR LUMBAR INTERBODY FUSION (PLIF) WAS PERFORMED AT L5/S. INTRA-OP, CAGE WAS PLACED ON THE LEFT SIDE, AND WHEN AN ATTEMPT WAS MADE TO INSERT THE SECOND CAGE ON THE RIGHT SIDE THE SECOND CAGE BROKE. THE SECOND CAGE WAS REPLACED WITH ANOTHER CAGE, AND WHEN AN ATTEMPT WAS MADE TO INSERT THAT CAGE, THAT CAGE WAS ALSO BROKE. BREAKAGE OCCURRED AT THE POSITION WHERE THE TIP OF THE CAGE WAS SLIGHTLY ENTERED TO THE INTERVERTEBRAL DISC SPACE (THE ANTERIOR MARKER WAS ABOUT AT THE POSTERIOR WALL OF THE VERTEBRAL BODY). THERE WERE NO FRAGMENTS REMAINED IN THE PATIENT. THE INTERVERTEBRAL DISC SPACE WAS NARROW AND MOBILITY WAS POOR. SHAVER AND DISTRACTOR WERE USED BEFORE INSERTION. THERE WAS A LESS THAN 60 MINUTES DELAY IN OVERALL PROCEDURE TIME AS A RESULT OF THIS EVENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10131975 ↗MalfunctionJun 9, 2020Break; Break
PRE-OPERATIVE DIAGNOSIS: KYPHOSIS PROCEDURE PERFORMED: POSTERIOR LUMBAR INTERBODY FUSION (PLIF) WAS PERFORMED AT L5/S. INTRA-OP, CAGE WAS PLACED ON THE LEFT SIDE, AND WHEN AN ATTEMPT WAS MADE TO INSERT THE SECOND CAGE ON THE RIGHT SIDE THE SECOND CAGE BROKE. THE SECOND CAGE WAS REPLACED WITH ANOTHER CAGE, AND WHEN AN ATTEMPT WAS MADE TO INSERT THAT CAGE, THAT CAGE WAS ALSO BROKE. BREAKAGE OCCURRED AT THE POSITION WHERE THE TIP OF THE CAGE WAS SLIGHTLY ENTERED TO THE INTERVERTEBRAL DISC SPACE (THE ANTERIOR MARKER WAS ABOUT AT THE POSTERIOR WALL OF THE VERTEBRAL BODY). THERE WERE NO FRAGMENTS REMAINED IN THE PATIENT. THE INTERVERTEBRAL DISC SPACE WAS NARROW AND MOBILITY WAS POOR. SHAVER AND DISTRACTOR WERE USED BEFORE INSERTION. THERE WAS A LESS THAN 60 MINUTES DELAY IN OVERALL PROCEDURE TIME AS A RESULT OF THIS EVENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10131968 ↗MalfunctionSep 20, 2018Crack; Crack
IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH LUMBAR CANAL STENOSIS AND UNDERWENT TRANSFORAMINAL LUMBAR INTERBODY FUSION. DURING THE SURGERY, THE CAGE CRACKED. AS THE INTER-VERTEBRAL DISC SPACE WAS NARROW, THE CAGE WAS INSERTED DURING DISTRACTION THROUGH THE EXTENDER. ADDITIONAL INSERTION WAS PERFORMED USING BONE IMPACTOR AND PLASTIC HAMMER. WHEN THE POSTERIOR SIDE OF THE CAGE WAS OBSERVED, A CRACK WAS FOUND. THE CAGE REMAINS IMPLANTED IN THE PATIENT. NO HEALTH DAMAGE TO THE PATIENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7896482 ↗MalfunctionMay 30, 2018Break; Positioning Problem; Break; Positioning Problem
PRE-OPERATIVE DIAGNOSIS FOR THIS PROCEDURE : LCS TYPE OF PROCEDURE OR TECHNIQUE USED: TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) LEVELS: L5/S1 IT WAS REPORTED THAT INTRA-OP, THE INTERVERTEBRAL DISC SPACE WAS MANAGED TO BE WIDENED FOR INSERTING THE CAGE. THE CAGE WAS INSERTED IN A WRONG DIRECTION. IT WAS HARD TO INSERT THE CAGE AND THE CAGE WAS BROKEN IN THE PROCESS OF BEING HIT. AFTER THAT, THE DIRECTION WAS CONFIRMED BY THE PERSPECTIVE IMAGE. IT WAS SAID THAT THERE WAS NO PROBLEM BECAUSE THE CAGE WAS INSERTED WITHOUT PROBLEMS AFTER THE IMAGE CONFIRMATION. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7554987 ↗InjuryFeb 7, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, ON (B)(6) 2016, PATIENT PRESENTED WITH PRE-OPERATIVE DIAGNOSIS: TH12 BURST FRACTURE AND LUMBAR SPINAL CANAL STENOSIS AND UNDERWENT PROCEDURE: TH10/L3: POSTERIOR SPINAL FUSION, L2/3: POSTERIOR LUMBAR INTERBODY FUSION. ON (B)(6) 2017, POST-OP, INFECTION AT PLIF CAGE WAS OBSERVED. PATIENT DID NOT ACHIEVE SOLID FUSION DUE TO INFECTION. ON (B)(6) 2017, ANTERIOR FUSION(REMOVAL OF CAGE, BONE AUTOGRAFTING) WAS CONDUCTED.
Patient problems: No Code Available; Unspecified Infection
MAUDE report 6306314 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page