CLYDESDALE PTC™ Spinal System 4986840Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CAGE 4986840 CDALE PTC 18MM 6 DEG 8X40 · Class 2 · Implant
All models of CLYDESDALE PTC Spinal System (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169190542), received between Aug 22, 2017 and Oct 8, 2024 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Material Deformation1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
- Neurological Deficit/Dysfunction1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Use of Device Problem; Use of Device Problem
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT PATIENT UNDERGOING OLIF OBLIQUE-LATERAL LUMBAR INTERBODY FUSION THERAPY FOR ADULT SPINAL DEFORMITY. LEVEL AT WHICH THE IMPLANT WAS PERFORMED AT T12-L5 IT WAS REPORTED THAT HERE WAS AN ERROR IN THE LEVEL L2/3 DURING THE OLIF IMPLEMENTATION PROCESS, SO A CAGE WAS INSTALLED IN L1/2 WHERE THE PROCEDURE HAD ALREADY BEEN PERFORMED, AND THE PREVIOUSLY INSTALLED CAGE DISLODGED TO CONTRALATERAL SIDE. THERE WAS LESS THAN 60 MINUTES DELAY IN THE OVERALL PROCEDURE TIME. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE CAGE THAT WAS ORIGINALLY INSTALLED IN L1/2 WAS PUSHED TO THE OPPOSITE SIDE. A CAGE WAS INSTALLED IN L2/3 AND THE WOUND WAS CLOSED. THE CAGE THAT DEVIATED TO THE OPPOSITE SIDE WAS REMOVED BY THE RIGHT APPROACH. THERE WAS NO INITIAL DEFECT WITH THE PRODUCT . ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THERE IS NO MALFUNCTION OR ALLEGATION ASSOCIATED WITH O-ARM OR ANY NAVIGATION DEVICE. IT IS THOUGHT THE THE WRONG LEVEL INSTALLATION WAS DUE TO THE USER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20397216 ↗MalfunctionOct 20, 2021Break; Material Deformation; Break; Material Deformation
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING PATIENT SPINAL THERAPY FOR SCOLI OSIS. IT WAS REPORTED THAT AFTER PLACING THE DD, THEY TRIED TO REMOVE IT BECAUSE IT WAS A LITTLE DEEPER POSTERIOR, BUT THE PART CONNECTING CAGE AND THE REMOVER WAS BROKEN AND THEY COULD ONLY REMOVE THE PRODUCT HALFWAY, SO THEY HAD NO CHOICE BUT TO INSERT THE PRODUCT AGAIN AS IT WAS. PROCEDURE INVOLVED IN THE EVENT WAS OBLIQUE LATERAL INTERBODY FUSION (OLIF) AND LEVELS IMPLANTED WAS L1/2.THERE WAS PATIENT INVOLVED IN THE EVENT AND NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT CAGE'S INSTRUMENT CONNECTION PART WAS DEFORMED, THERE IS ONLY THE POSSIBILITY OF BREAKAGE. THE REPORTED PRODUCTS CAME IN CONTACT WITH THE PATIENT. THE PRODUCTS ARE UNDER IMPLANTATION AND NO PLAN TO RETURN THE PRODUCTS.
Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions
MAUDE report 12664152 ↗InjuryDec 22, 2017Break; Break
PRE-OPERATIVE DIAGNOSIS FOR THIS PROCEDURE: LUMBAR DEGENERATIVE SCOLIOSIS TYPE OF PROCEDURE BEING PERFORMED : POSTERIOR DECOMPRESSION FUSION, OBLIQUE LUMBAR INTERBODY FUSION LEVELS: L2/3 IT WAS REPORTED THAT INTRA-OP, O-ARM IMAGE REVEALED THAT THE CAGE POSITION WAS IN THE SPINAL CANAL BEFORE CLOSURE. IT WAS REMOVED IMMEDIATELY, AND THE INCISION WAS CLOSED. THE CAGE BROKE. AGGRAVATION OF NEUROLOGICAL SYMPTOMS IN THE PATIENT WAS OBSERVED NEXT DAY.
Patient problems: Neurological Deficit/Dysfunction
MAUDE report 7141652 ↗MalfunctionOct 26, 2017Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT OBLIQUE LUMBAR INTERBODY FUSION (OLIF) AT L3-5 DUE TO LUMBAR DEGENERATIVE SCOLIOSIS. DURING INSERTION, THE CAGE BROKE. THE CAGE WAS INSERTED TO INSERTER FOR L3/4, IT COULD NOT BE REMOVED FROM THE INSERTER THEREAFTER. THE SURGEON TRIED TO REMOVE IT FORCIBLY AND THE EDGE OF THE CAGE BROKE. THE FRAGMENTS WERE SECURED AND OTHER THAN THAT IT REMAINED IN THE PATIENT. WHEN THE SAME INSERTER WAS USED FOR L4/5 TO INSERT THE CAGE, THERE WAS NO TROUBLE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6978959 ↗MalfunctionAug 22, 2017Break; Break
IT WAS REPORTED THAT PATIENT WITH PREOPERATIVE DIAGNOSIS OF SPINAL CANAL STENOSIS AND KYPHOSIS UNDERWENT OBLIQUE LUMBAR INTERBODY FUSION (OLIF). INTRA-OP, WHEN THE SPINAL CAGE WAS TRIED TO BE INSERTED IN L1/2 USING INSERTER, IT BROKE NEAR THE ENTRANCE. SINCE THE INTERVERTEBRAL SEGMENT WAS VERY TIGHT, THE CAGE WAS HAMMERED WITH SLIGHTLY STRONG FORCE. THE INTERVERTEBRAL DISC WAS SO NARROWED THAT A COBB COULD NOT BE INSTALLED ON THE OTHER SIDE. AS RIBS AND PERITONEUM DISTURBED INSERTION OF CAGE, THE CAGE COULD NOT BE INSERTED PARALLEL TO THE INTERVERTEBRAL DISC. THE SURGEON INSERTED CAGE FROM THE CAUDAL SIDE TO THE CRANIAL SIDE, AND CAGE WAS BROKEN. THIS WAS EXPLANTED. EVEN TRIAL WAS NOT CARRIED OUT. THEN, NEW CAGE WAS PREPARED. IN THIS TIME, TRIAL WAS CONDUCTED AND THE TAB OF THE INSERTER WAS INSERTED ALL THE WAY. AS A RESULT, THE NEW ONE COULD BE INSERTED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6809421 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.