CLYDESDALE PTC™ Spinal System 4986255Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CAGE 4986255 CDALE PTC 18MM 6 DEG 12X55 · Class 2 · Implant
All models of CLYDESDALE PTC Spinal System (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169190665), received between Feb 25, 2019 and Sep 6, 2022 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 6, 2022Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH MEDICAL HISTORY OF DM. PREOPERATIVE DIAGNOSIS OF THIS SURGERY WAS L2-5. LCS. PROCEDURE OR THERAPY PERFORMED WAS OLIF. LEVEL AT WHICH THE IMPLANT WAS PERFORMED WAS L2-5. IT WAS REPORTED THAT AN OLIF WAS PERFORMED IN ORDER FROM L2/3 HEAD SIDE. AFTER PERFORMING IN THE L3/4, WHEN IT WAS EXPANDED IN L4/5, A PIECE LIKE PLASTIC WAS CONFIRMED IN THE SURGICAL FIELD. THE ENTIRE CAGE WAS INSERTED IN THE BODY, SO IT COULD NOT BE CONFIRMED BUT IT SEEMS LIKE A FRAGMENT OF THE CAGE INCLUDING ITS COLOR. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15359111 ↗MalfunctionSep 6, 2022Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH MEDICAL HISTORY OF DM. PREOPERATIVE DIAGNOSIS OF THIS SURGERY WAS L2-5. LCS. PROCEDURE OR THERAPY PERFORMED WAS OLIF. LEVEL AT WHICH THE IMPLANT WAS PERFORMED WAS L2-5. IT WAS REPORTED THAT AN OLIF WAS PERFORMED IN ORDER FROM L2/3 HEAD SIDE. AFTER PERFORMING IN THE L3/4, WHEN IT WAS EXPANDED IN L4/5, A PIECE LIKE PLASTIC WAS CONFIRMED IN THE SURGICAL FIELD. THE ENTIRE CAGE WAS INSERTED IN THE BODY, SO IT COULD NOT BE CONFIRMED BUT IT SEEMS LIKE A FRAGMENT OF THE CAGE INCLUDING ITS COLOR. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15359091 ↗InjuryFeb 27, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT OBLIQUE LATERAL INTERBODY FUSION (OLIF) SURGERY AT L3/4/5 DUE TO DISH (DIFFUSE IDIOPATHIC SKELETAL HYPEROSTOSIS) PATIENT AND RESTENOSIS AFTER POSTERIOR DECOMPRESSION. TWO WEEKS LATER, THE PATIENT WAS HOSPITALIZED FOR THE SCHEDULE OF POSTERIOR FIXATION, BUT THE PATIENT COMPLAINED OF PAIN AND WAS DIAGNOSED BY MAGNETIC RESONANCE IMAGING (MRI). CLEANING WAS PERFORMED DUE TO THE SUSPECTED INFECTION. WHEN THE CAGE OF L3/4 WAS TOUCHED, IT WAS FOUND TO BE UNSTEADY, SO REMOVAL WAS ALSO PERFORMED. CLEANING WAS PERFORMED REPEATEDLY AND THE WOUND WAS CLOSED. IT SEEMED THAT ICU MANAGEMENT WAS NECESSARY AFTER SURGERY AND THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL WHILE BEING INTUBATED.
Patient problems: Pain; Unspecified Infection
MAUDE report 9761391 ↗InjuryFeb 25, 2019Break; Break
PRE-OPERATIVE DIAGNOSIS: SPONDYLOLISTHESIS PROCEDURE: OBLIQUE LUMBAR INTERBODY FUSION (OLIF) LEVEL IMPLANTED: L4/5 IT WAS REPORTED THAT INTRA-OP, THE PHYSICIAN PREPARED A CAGE OF A BIGGER SIZE AND INSERTED THE BIGGER CAGE INSIDE THE PATIENT'S BODY, BUT BECAUSE THERE WAS CAGE IN THE SAME POSITION AS THE POSITION WHERE INSTALLED THE ORIGINAL CAGE, THE BIGGER CAGE WAS REMOVED TOO. THE THREAD PART OF THE BIGGER CAGE BROKE. ADDITIONALLY, SURGERY PERFORMED BECAUSE CAGE INSERTION BY OLIF APPROACHING FAILED SO POSTERIOR LUMBAR INTERBODY FUSION WITH POSTERIOR APPROACHING WAS PERFORMED. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT SYMPTOMS OR COMPLICATION WERE REPORTED AS RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8367904 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗