CLYDESDALE PTC™ Spinal System 4986245Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CAGE 4986245 CDALE PTC 18MM 6 DEG 12X45 · Class 2 · Implant
All models of CLYDESDALE PTC Spinal System (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169190689), received between Mar 17, 2017 and Jun 22, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Break1
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 22, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT OBLIQUE LUMBAR INTERBODY FUSION - OLIF AT L3/4, L4/5. INTRA-OP, DURING HAMMERING WITH GRASPING THE SHAFT PART BELOW THE NUT OF THE INSERTER, A GLOVE WAS CAUGHT IN, AND THE SHAFT COULD NOT MOVE. THE INSERTER WAS REMOVED ONCE, AND THE CAGE WAS REMOVED, BUT THE SHAFT STILL COULD NOT MOVE. ANOTHER INSERTER WAS USED, BUT BECAUSE THE THREAD OF THE CAGE WAS BROKEN AND IT COULD NOT BE GRASPED, A NEW ONE WAS UNPACKED. A SCRUB NURSE MADE THE SHAFT MOVE WITH USING A REMOVER AND THE GLOVE WHICH WAS CAUGHT IN WAS REMOVED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7628301 ↗MalfunctionMar 17, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT PATIENT WITH THORACIC OSSIFICATION OF THE YELLOW LIGAMENT AND OSSIFICATION OF POSTERIOR LONGITUDINAL LIGAMENT UNDERWENT OBLIQUE LUMBAR INTERBODY FUSION AT L2/3. INTRA-OP, WHEN SURGEON WAS INSERTING CAGE AT L2/3 USING INSERTER, THE CAGE CAME OFF FROM THE INSERTER. THE SURGEON COULDN'T ADD MORE IMPACT TO THE CAGE, SO THE CAGE WAS REMOVED USING FORCEPS. EXCESSIVE FORCE WAS APPLIED ON INSERTER DURING CAGE INSERTION. THREADS OF THE CAGE WERE SHAVED.THE SURGEON TRIED TO REINSERT THE CAGE BUT COULDN'T INSERT IT. ANOTHER CAGE WAS OPENED AND USED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6412507 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page