Skip to content

CLYDESDALE PTC™ Spinal System 4922255Recalls & reported problems

Medtronic Sofamor Danek, Inc. · CAGE 4922255 CDALE PTC 22MM 12 DEG 12X55 · Class 2 · Implant

All models of CLYDESDALE PTC Spinal System (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169191310), received between Oct 9, 2018 and Apr 18, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2024Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DIRECT LATERAL INTERBODY FUSION (DLIF) APPROACH. IT WAS REPORTED THAT THE DEVICE WAS DAMAGED. WHILE THE SURGEON WAS IMPLANTING THE PRODUCT, WE NOTICED THE INSERTER/IMPLANT MADE A DISTINCT NOISE. AFTER LOOKING AT THE XRAY, WE NOTICE THE INSERTER WAS BENT AND THE IMPLANT WAS DAMAGED. SO, WE ARE NOT SURE IF THE IMPLANT CAUSED THE INSERTER TO BEND OR IF THE INSERTER BENDING CAUSED THE IMPLANT TO BREAK. THE MALLETS WERE BROKEN. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT IT WAS A FAILURE ON THE INSERTER PRONGS THAT CREATED THE BREAKAGE/FAILURE ON THE IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19132219 ↗
  • MalfunctionOct 9, 2018Break; Break

    PRE-OPERATIVE DIAGNOSIS FOR THIS PROCEDURE: LUMBAR SPINAL CANAL STENOSIS TYPE OF PROCEDURE OR TECHNIQUE USED: ANTERIOR LUMBAR FUSION AND POSTERIOR LUMBAR FUSION IT WAS REPORTED THAT INTRA-OP, THREAD OF THE CAGE WAS DEFORMED OR BROKEN. INTRA-OPERATIVELY, AFTER CLEANING UP THE INTERVERTEBRAL DISC, WHEN ABOUT 90% OF THE CAGE WAS INSERTED AT L4-5, IT WAS DIFFICULT TO INSERT THE CAGE DEEPER. THE PHYSICIAN TRIED TO HIT THE INSERTER TO CONTRALATERAL SIDE WITH A HAMMER IN ORDER TO REMOVE THE CAGE, HOWEVER IT COULD NOT BE REMOVED. THE CAGE WAS THEN REMOVED BY REMOVAL IMPLEMENT AND THE INTERVERTEBRAL DISC WAS CURETTED AGAIN. WHEN TRYING TO ATTACH THE CAGE TO THE INSERTER AGAIN, THE CAGE COULD NOT BE GRIPPED DUE TO PROBABLE BREAKAGE OF THREAD OF THE CAGE. A NEW CAGE WAS THEN OPENED TO DEAL WITH THIS PROBLEM. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7948896 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates