NC SPRINTER® NCSP37515XRecalls & reported problems
Medtronic, Inc. · BALLOON NCSP37515X NCSPRINTER RX M US · Class 2
All models of NC SPRINTER (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169282070), received between Feb 23, 2026 and Feb 23, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
- Improper or Incorrect Procedure or Method1
- Material Deformation1
Patient problems
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 23, 2026Burst Container or Vessel; Improper or Incorrect Procedure or Method; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING ONE NC SPRINTER RX BALLOON CATHETER, THE BALLOON BURST DURING BALLOON INFLATION AT 20ATM, EXCEEDING THE RATED BURST PRESSURE (RBP) OF 18 ATM. THE TARGET LESION WAS LOCATED IN THE PROXIMAL CIRCUMFLEX (CX) ARTERY WHICH EXHIBITED NO TORTUOSITY AND MODERATE CALCIFICATION. THE DEVICE WAS NOT INSPECTED PRIOR TO USE AND NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED. THE DEVICE WAS NOT KINKED AND RESTRAIGHTENED DURING USE. THE DEVICE DID PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. THE PROXIMAL CIRCUMFLEX LESION WAS PRE-DILATED WITH A 2.5MM NON-COMPLIANT (NC) BALLOON, FOLLOWED BY A 3.0MM NC BALLOON, FOLLOWED BY A 3.5 MM SHOCKWAVE BALLOON. A GUIDELINER WAS THEN USED TO DELIVER A 3.5 X 26 RESOLUTE INTEGRITY STENT. POST-IMPLANT OF THE STENT, A 4.5 MM NC BALLOON WAS USED TO DILATE THE DISTAL AND MID SECTION OF THE STENT. THE 3.75 X 15 NC SPRINTER BALLOON WAS SMOOTHLY DELIVERED TO THE PROXIMAL END OF THE STENT THROUGH THE GUIDELINER AND INFLATED TO 20 ATM. THE BALLOON RUPTURED. THE BALLOON CATHETER WAS WITHDRAWN FRO
Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information
MAUDE report 24412986 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗