NC SPRINTER® NCSP3015XRecalls & reported problems
Medtronic, Inc. · BALLOON NCSP3015X NCSPRINTER RX M US · Class 2
All models of NC SPRINTER (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169282421), received between Dec 13, 2022 and Dec 13, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
- Device Dislodged or Dislocated1
Patient problems
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 13, 2022Burst Container or Vessel; Detachment of Device or Device Component; Device Dislodged or Dislocated; Burst Container or Vessel; Detachment of Device or Device Component; Device Dislodged or Dislocated
AN ATTEMPT WAS MADE TO USE ONE NC SPRINTER PTCA BALLOON CATHETER TO TREAT A NON-TORTUOUS, NON-CALCIFIED LESION WITH 70% STENOSIS, IN THE DISTAL LEFT MAIN (LM) CORONARY ARTERY AND LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE LEFT ANTERIOR DESCENDING (LAD) ARTERY HAD 70% IN-STENT RESTENOSIS. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. A 3.5 NON-MEDTRONIC GUIDEWIRE WAS PLACED IN THE DISTAL LAD. THE LESION WAS PRE-DILATED WITH A 2.0 X 10MM BALLOON AND A 3.0 X 12MM SPRINTER LEGEND BALLOON UP TO 16 ATM. A 3.0 X 15MM RESOLUTE INTEGRITY DRUG ELUTING STENT WAS IMPLANTED AT THE PROXIMAL LAD AT 18 ATM. THE STENT WAS BEING POST-DILATED WITH THE NC SPRINTER BALLOON AT 16 ATM. IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED DURING THE BALLOON INFLATION AT 16 ATM. THE RUPTURED BALLOON WAS RETAINED IN THE LAD. AN ATTEMPT WAS MADE TO REMOVE THE RUPTURED BALLOON WITH NON-MEDTRONIC GUIDE EXTENSION CATHETERS AND A 4.0 SNARE, BUT IT WAS NOT POSSIBLE TO REMOVE. A 3.0 X 22MM RESOLUTE INTEGRITY STENT WAS USED TO ENTRAP THE REMNANT BALLOON AT THE PROXIMAL LAD AT 18 ATM. FINAL ANGIOGRAM SHOWED GOOD RESULT WITH TIMI 3 FLOW. IT WAS STATED THAT THE STENT WAS INSERTE
Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information
MAUDE report 15967924 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.