NIM-Response® 8253200 (UDI 00643169283275)Recalls & reported problems
Medtronic Xomed, Inc. · PATIENT INTFC 8253200 RESPONSE 3.0 ROHS · Class 2
All models of NIM-Response (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Xomed NIM-Response patient interface is a nerve stimulator accessory. Its 99 adverse event reports include no deaths, 1 injury and 98 malfunctions. The reports mostly describe the interface giving no stimulation response before or during surgery, and unclear or incorrect readings. Most reports describe no clinical signs or patient impact. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 100
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 00643169283275), received between Dec 27, 2016 and Sep 17, 2026 — classified as 1 injury, 99 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (33 reports).
3 other recalls cover devices cleared under the same FDA submission (K083124); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings33
- Use of Device Problem30
- False Negative Result9
- Device Sensing Problem8
- Inappropriate Audible Prompt/Feedback6
- Break4
- Failure to Sense4
- Device Operates Differently Than Expected3
Patient problems
- No Clinical Signs, Symptoms or Conditions73
- No Consequences Or Impact To Patient23
- No Known Impact Or Consequence To Patient7
- No Patient Involvement2
- Injury1
Recent adverse event reports
The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Inappropriate Audible Prompt/Feedback
It was reported that pre op there was a great interference occurred from the patient interface. There was no patient involved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50798285 ↗MalfunctionJun 15, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING SET-UP, THE PATIENT INTERFACE HAD NO RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25498386 ↗MalfunctionJun 13, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PRE-OP, NIM SYSTEM HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25494700 ↗MalfunctionMay 8, 2026False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING PRE-OP, THE NIM 3.0 HAD NO STIM RESPONSE. THERE WAS RETURN CURRENT. ISSUE NOT RESOLVED BY REPOSITIONING OR TROUBLESHOOTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25120157 ↗MalfunctionMay 7, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING PRE-OP, THERE WAS NO STIMULATION RESPONSE IN PATIENT INTERFACE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25108493 ↗MalfunctionMar 23, 2026Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT DURING OPERATION THE PI BOX WHEN PLUGGING IN IT DOES NOT RESPOND. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24659228 ↗MalfunctionFeb 13, 2026False Negative Result; Connection Problem
IT WAS REPORTED THAT PRE-OP, NIM PATIENT INTERFACE HAD NO STIM RESPONSE. POOR CONTACT AT STIM PORT, NO IRRITATION REACTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24330280 ↗MalfunctionDec 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, IT WAS FOUND THAT THERE IS NO STIMULATION RESPONSE ( USED THE SIMULATOR TEST MACHINE BEFORE OPERATION , CONTACT SIGNAL IS WEAK ). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23815774 ↗MalfunctionDec 3, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING THE OPERATION, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THE STIM RETURN, LOCATED UNDER THE STIMULUS SETTING ON THE MONITOR, SHOWS 0. THERE IS NO STIMULUS DELIVERY TONE HEARD WHEN ATTEMPTING TO STIMULATE THE NERVE. THERE IS NO WAVEFORM DISPLAYED ON THE MONITOR WHEN ATTEMPTING TO STIMULATE THE NERVE. THERE WAS NO PATIENT IMPACT RELATED TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23712370 ↗MalfunctionNov 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE ELECTRODE BOX WAS DEFECTIVE, THE MAIN DEVICE ITSELF WAS WORKING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23663556 ↗MalfunctionSep 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PRE-OPERATION, THE PATIENT INTERFACE WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22994859 ↗MalfunctionAug 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, IT WAS FOUND THAT THE DEVICE DOES NOT STIMULATE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22892737 ↗MalfunctionMay 9, 2025False Negative Result; False Negative Result
IT WAS REPORTED THAT PRE-OP, NIM PATIENT INTERFACE HAD NO STIMULATION RESPONSES. STIM1 CHANNEL SELF-TEST CAN PASS, BUT THERE IS NO STIMULUS RESPONSE. THE PROCEDURE WAS COMPLETED USING BACK UP DEVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21991391 ↗MalfunctionMar 14, 2025False Negative Result; False Negative Result
IT WAS REPORTED THAT BEFORE OPERATION, THERE WAS NO SIGNAL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21603801 ↗MalfunctionMar 10, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING A PAROTID OPERATION, THERE WAS A DELAYED CURRENT DELIVERY TONE IN THE NIM 3.0 SYSTEM. THE FOLLOWING STEPS WERE TAKEN: THE MUTING DETECTOR WAS TURNED OFF, THE VOLUME WAS INCREASED, THE THRESHOLD WAS LOWERED, THE PATIENT WAS NOT PARALYZED, THE ELECTRODE CHECK WAS PASSED, THE TAP TEST WAS PASSED, THE PROBE WAS EXCHANGED, THE STIMULATION WAS INCREASED, THE FUSE IN THE PI BOX WAS NOT BLOWN, AND THE ELECTRODES WERE CHANGED. DESPITE THESE STEPS, THE DELIVERY TONE WAS STILL DELAYED. THE SURGERY WAS EXTENDED BY LESS THAN ONE HOUR. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21561189 ↗MalfunctionMar 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT PRIOR TO PROCEDURE NIM PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21543662 ↗MalfunctionMar 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Use of Device Problem; Incorrect, Inadequate or Imprecise Result or Readings; Use of Device Problem
IT WAS REPORTED THAT THE THERE WAS NO STIMULATION RESPONSE AS STIM 1 WAS NOT WORKING IN THE PATIENT INTERFACE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21536379 ↗MalfunctionMar 5, 2025False Negative Result; False Negative Result
IT WAS REPORTED THAT PRIOR TO USE, THE DEVICE WAS NON RESPONSIVE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21525057 ↗MalfunctionFeb 26, 2025False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT DURING INTRA-OP, THE ELECTRODE WAS NOT SHOWING UP WITH THE PATIENT INTERFACE. IT JUST MAKE NOISE WHEN PLUGGED IN WITH NO SIM ON IT. SWAPPED WITH ANOTHER NIM UNIT AND THE ISSUE WAS RESOLVED. THERE WAS NO KNOWN PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21470522 ↗MalfunctionFeb 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED DURING PRE-OP, THAT THE DEVICE HAD CHANNEL BLOCKAGE, NO SIGNAL. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21466638 ↗MalfunctionFeb 25, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE DEVICE DOES NOT SHOW THE STIMULATION RESPONSE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21456436 ↗MalfunctionJan 23, 2025Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback; Incorrect, Inadequate or Imprecise Result or Readings; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT NIM GENERATOR DOES NOT WORK DURING PATIENT TESTING. NO MORE TRACE AND SPECIFIC SOUND BUT JUST A BEEP WHEN POSITIONING THE NIM IN CONTACT WITH THE NERVES. CHANGE OF NIM DEVICE, INTUBATION TUBE. CHECKING THE CONNECTIONS. PROBLEM STILL EXISTING. THERE WAS NO PATIENT IMPACT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21214888 ↗MalfunctionNov 13, 2024Inappropriate Audible Prompt/Feedback; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT PRIOR TO PROCEDURE NIM PATIENT INTERFACE HAD INTERFERENCE. THE PROCEDURE WAS COMPLETED WITH THE BACK UP DEVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20671057 ↗MalfunctionNov 1, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE THE OPERATION, THE PATIENT INTERFACE HAD NO STIMULATION RESPONSE. THERE IS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20593829 ↗MalfunctionOct 17, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT BEFORE OPERATION, THERE WAS NO STIMULATION RESPONSE. THERE WAS NO PATIENT INVOLVED IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20468321 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083124. One submission often covers many models, so check the product and lot details in each record.
- Z-2344-2012 · Sep 27, 2010 · TerminatedInterfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253210 (Pulse). LABELING: Box Label: REF 8253210 x1***NIM-Response 3.0 Patie
- Z-2342-2012 · Sep 27, 2010 · TerminatedInterfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces 8253200 (Response) LABELING: Box Label: REF 8253200 x1***NIM-Response 3.0 Patient
- Z-2343-2012 · Sep 27, 2010 · TerminatedInterfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253410 (Neuro). LABELING: Box Label: REF 8253410 x1***NIM-Response 3.0 Patie
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page