Midas Rex Clearview™ SP12MH30Recalls & reported problems
Medtronic Xomed, Inc. · TOOL DC SP12MH30 LGD 12CM PROX 3.0 MH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169308497), received between Jul 18, 2019 and Sep 16, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K081277); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation4
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
- Radiation Exposure, Unintended1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2021Material Separation; Material Separation
ADDITIONAL INFORMATION RECEIVED MENTIONED THAT THERE WERE NO FRAGMENTS THAT GOT DETACHED OR CAME OFF FROM THE BUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12480866 ↗MalfunctionSep 24, 2020Material Separation; Material Separation
IT WAS REPORTED THAT WHEN THE SURGEON WAS USING THE DRILL BIT, IT BROKE. THE PARTS OF THE BROKEN DRILL BIT WERE ALL COLLECTED. THERE WAS NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. THE INTENDED PROCEDURES WERE LUMBAR 4-LUMBAR 5 AND LUBAR 5 SCAL ONE HEMILAMINECTOMIES, MEDIAL FACETECTOMIES, FORAMINOTOMIES, POSTERIOR INSTRUMENTED FUSION, POSSIBLE INTERBODY FUSION, ALLOGRAFT, AUTO GRAFT, POSSIBLE BONE MORPHOGENIC PROTEIN. UPON FOLLOW UP IT WAS CONFIRMED THAT ANOTHER DRILL BIT WAS USED IN THE SURGERY, THERE WAS NO SIGNIFICANT DELAY, THE SURGEON WAS USING FLUOROSCOPE, THAT AID THE DOCTOR IN GETTING ALL OF THE FRAGMENTS, USING FORCEPS (TWEEZERS). TO CONFIRM, THE SURGICAL TECH IN THE OR, COMPARED THE PIECES OF THE BROKEN DRILL BIT THEY HAD TO THE WHOLE DRILL BIT USED TO COMPLETE THE SURGERY, AND BELIEVED THEY RETRIEVED ALL OF THE PIECES/ FRAGMENTS. THE DEVICE WAS IN THE CONTAINER AND LOOKED 2 GOOD SIZE PIECES BROKE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10579836 ↗MalfunctionFeb 17, 2020Material Separation; Material Separation
A HEALTHCARE PROVIDER REPORTED VIA MANUFACTURER REPRESENTATIVE THAT THE TIP OF THE BUR HAD BROKEN OFF/SHEARED OFF INTO THE PATIENT. THE SURGEON WAS USING THE BUR WITH IRRIGATION, DISSECTING BONE ON THE VERTEBRAL. WHEN THE DRILL WAS REMOVED, THE TIP OF THE BUR WAS SITTING ON THE VERTEBRAL BODY. THIS WAS REMOVED WITH A FORCEP AND A NEW BUR WAS OPENED. THERE WERE NO BROKEN PIECES THAT REMAINED INSIDE THE PATIENT'S BODY. THE BUR HAD NOT BEEN USED FOR AN EXTENSIVE AMOUNT OF TIME, IT WAS BEING USED WITH IRRIGATION AND ALSO WASN'T HAVING EXCESSIVE FORCE APPLIED TO IT. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9716572 ↗InjuryJul 18, 2019Material Separation; Material Separation
A HEALTH CARE PROVIDER (HCP) REPORTED THAT A DEVICE TIP BROKE OFF IN PATIENT BACK. X-RAY WAS TAKEN AND ALL PIECES WERE RETRIEVED.
Patient problems: Radiation Exposure, Unintended
MAUDE report 8803138 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081277. One submission often covers many models, so check the product and lot details in each record.
- Z-0961-2022 · Dec 15, 2021 · TerminatedSoftware 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND EH
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page