Skip to content

ZEVO™ Anterior Cervical Plate System G3001019Recalls & reported problems

Medtronic Sofamor Danek, Inc. · PLATE G3001019 ZEVO 19MM 1 LVL · Class 2 · Implant

All models of ZEVO Anterior Cervical Plate System (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169327382), received between Jan 1, 2020 and Jan 1, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

1 other recall cover devices cleared under the same FDA submission (K141632); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 1, 2020Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT CERVICAL ANTERIOR FIXATION AT C6/7 DUE TO CERVICAL SPINE TRAUMA. POST-OP, THE SCREWS WERE INTEGRATED WITH TWO PLATES AND IT SEEMED THAT THE LOWER PART OF THE PLATE HAS COME OF RATHER THAN THE SCREW CAME OFF AND HAS BEEN A LITTLE UNSTABLE. THERE WAS AN IMPRESSION THAT INTRA-OPERATIVE, BONE QUALITY WAS QUITE BAD. HENCE ON (B)(6) 2019, FIXATION WAS PERFORMED FROM THE POSTERIOR SIDE(VS WAS USED AT C4/5/6, T1/2/3) BUT REMOVAL HAD NOT BEEN PERFORMED. THERE WERE NO OTHER PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9537115 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141632. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2018 · Mar 23, 2018 · Terminated
    Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page