Skip to content

ZEVO™ Anterior Cervical Plate System G3002037Recalls & reported problems

Medtronic Sofamor Danek, Inc. · PLATE G3002037 ZEVO 37MM 2 LVL · Class 2 · Implant

All models of ZEVO Anterior Cervical Plate System (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169327504), received between Jan 3, 2025 and Jan 3, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

1 other recall cover devices cleared under the same FDA submission (K141632); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2025Migration; Migration

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTERIOR CERVICAL FUSION FOR CERVICAL SPONDYLOTIC MYELOPATHY. IT WAS REPORTED THAT ON JULY 7, THE ZV WASINITIALLY USED. LESS THAN 3 WEEKS AFTER THE SURGERY, DISLODGEMENT (MIGRATION) OF THE BONE GRAFT WAS OBSERVED. COMPARATIVE IMAGES OF POSTOPERATIVE AND MOST RECENT X-RAY IMAGING WERE OBTAINED. THE GRAFTED BONE COLLAPSED 20¿30° POSTERIORLY. AS A RESULT, THE L4 LOWERENDPLATE - THE L6 UPPER ENDPLATE WILL DECREASE BY 4-5 MM, AND THE C6 SCREW ANGLE WILL BE -10° (IT WAS BELIEVED THAT THE VARIABLE MECHANISM WAS ACTIVATED). THE PATIENT RECEIVED AN ILIAC BONE GRAFT FOR THE C5 CORPECTOMY AND PLATE FIXATION FOR THE C4-6. THE GENERAL FLOW OF THE PROCEDURE IS AS FOLLOWS: DEPLOYMENT WITH A PIN DISTRACTOR, ILIAC BONE GRAFTING, AND FIXATION WITH A PLATE. SINCE IT WASFIX-GROUP, THE GRAFTED BONE WAS MORE FIRM THAN LOOSE; IT WAS SLIGHTLY TAPPED AND INSERTED DURING THE ILIAC GRAFT. WHEN DISTRACTION WAS LOOSENED AND CHECKED, THERE WAS NO FEELING OF MOVEMENT, AND IT WAS STEADY. THE VERTEBRAL ENDPLATE WAS ALSO NOT DISRUPTED. NO SINKING WAS OBSERVED, AND IT SEEMED LIKE IT HAD DISLODGED BEHI

    Patient problems: Pain

    MAUDE report 21065363 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141632. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2018 · Mar 23, 2018 · Terminated
    Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page