ZEVO™ Anterior Cervical Plate System G3003055Recalls & reported problems
Medtronic Sofamor Danek, Inc. · PLATE G3003055 ZEVO 55MM 3 LVL · Class 2 · Implant
All models of ZEVO Anterior Cervical Plate System (39)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169327511), received between Aug 1, 2017 and Apr 11, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 other recall cover devices cleared under the same FDA submission (K141632); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Migration or Expulsion of Device1
Patient problems
- Discomfort1
- No Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 11, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING REVISION SURGERY. PRE-OP DIAGNOSIS WAS CERVICAL KYPHOSIS. IT WAS REPORTED THAT, PATIENT HAD DISCOMFORT DUE TO PLATE DISLODGEMENT. AFTER C5/6 DECOMPRESSION, 11MM CAGE WAS PLACED AND MIGRATION OF C5/6 CAGE ALREADY OCCURRED AFTER COMBINED POSTERIOR FIXATION. AFTER 1 WEEK POSTOPERATIVELY, FURTHER MIGRATION OF THE CAGE AND HEAD SIDE DISLODGEMENT OF THE C4-7 ZEVO PLATE OCCURRED. TOTAL REMOVAL OF PLATE AND REMOVAL OF 11MM CAGE AT C5/6, AUTOLOGOUS ILIAC BONE TRANSPLANTATION BETWEEN C5/6 VERTEBRAE WAS RETRIEVED DURING REVISION SURGERY. THE 13MM SCREW AT C4 AND 15MM SCREW AT C7 WAS FOUND LOOSE AND THEY WERE EXPLANTED. PROCEDURE OR TECHNIQUE PERFORMED DURING INITIAL SURGERY WAS C4-7 ANTERIOR CERVICAL DISCECTOMY AND FUSION. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: Discomfort
MAUDE report 19088968 ↗MalfunctionAug 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT PATIENT WITH CERVICAL SPONDYLOTIC MYELOPATHY UNDERWENT ANTERIOR CERVICAL DECOMPRESSION AND FUSION AT C3/6 AND CORPECTOMY AT C4/C5 AND THEN GRAFT BONE(AUTOGRAFT) WAS FILLED. POST-OP, IT WAS FOUND THAT THE PLATE BACKED OUT. AS X-RAY HAS NOT BEEN CONFIRMED, NEITHER SEVERITY NOR DETAIL IS KNOWN. IF BONE UNION IS ACHIEVED DURING FOLLOW-UP, RE-OPERATION WILL NOT BE CONDUCTED. PATIENT COMPLICATIONS HAVE NOT BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Information
MAUDE report 6757887 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141632. One submission often covers many models, so check the product and lot details in each record.
- Z-1750-2018 · Mar 23, 2018 · TerminatedMedtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page