ZEVO™ Anterior Cervical Plate System 7713518Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 7713518 ZEVO VAR SD 3.5MM X 18MM · Class 2 · Implant
All models of ZEVO Anterior Cervical Plate System (39)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169329300), received between Mar 14, 2019 and Jan 25, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K141632); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
Patient problems
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 25, 2025Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A SCREW USED IN ANTERIOR CERVICAL DISCECTOMY AND FUSION THERAPY. IT WAS REPORTED THAT THE SCREW HEAD BROKE OFF FROM THE SHAFT UPON INSERTION. THE SHANK OF THE SCREW REMAINED IN THE VERTEBRAL BODY, UNABLE TO BACK OUT ONCE THE HEAD OF THE SCREW SHEARED OFF FROM THE SHANK DURING IMPLANTATION. IT WAS A NORMAL APPROACH TO PLACING SCREWS. AWL WAS USED TO CREATE PILOT HOLE FOR THE SCREW. THIS WAS THE FIFTH SCREW GOING IN OF EIGHT TOTAL. THE HEAD OF THE SCREW SHEARED OFF FROM THE SHANK WHEN ADVANCING THE SCREW THROUGH THE SCREW HOLE IN THE PLATE AND INTO THE VERTEBRAL BODY. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 21236283 ↗MalfunctionMar 14, 2019Break; Break
PRE-OPERATIVE DIAGNOSIS: ANTERIOR CERVICAL DISC FUSION PROCEDURE: ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) LEVEL IMPLANTED: C4 IT WAS REPORTED THAT, INTRA-OP, SCREW CAUGHT PLATE, AND WHEN SCREW ADVANCED, IT SEEMED THE TORQUE CAUSED THE HEAD TO POP OFF THE SCREW. DUE TO THIS CAUSE SCREW HEAD DETACHED FROM THE SCREW. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. SCREW HEAD DETACHED FROM THE SCREW AND FRAGMENTS REMAINED IN THE PATIENT. NO PATIENT COMPLICATION WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8422042 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141632. One submission often covers many models, so check the product and lot details in each record.
- Z-1750-2018 · Mar 23, 2018 · TerminatedMedtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.