Skip to content

ZEVO™ Anterior Cervical Plate System 3002037Recalls & reported problems

Medtronic Sofamor Danek, Inc. · PLATE 3002037 ZEVO 37MM 2 LVL · Class 2 · Implant

All models of ZEVO Anterior Cervical Plate System (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169329966), received between Nov 19, 2024 and Nov 19, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

1 other recall cover devices cleared under the same FDA submission (K141632); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2024Migration; Migration

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PATIENT UNDERGONE SPINAL THERAPY. IT WAS REPORTED THAT SCREW AT C3 ON THE PATIENTS LEFT SIDE APPEARED TO HAVE BACKED OUT AFTER FLIPPING THE PATIENT FOR THE POSTERIOR PROCEDURE.THE SCREW HAD BACK OUT SLIGHTLY AND WAS OUTSIDE THE LOCK MECHANISM AND THE LOCK MECHANISM AT THE TOP AND MIDDLE OF THE PLATE WERE ONLY PARTIALLY LOCKED. THE SURGEON ADVANCED THE SCREW AND LOCKED THE MECHANISMS. PHYSICIANS COMMENTED THAT THE PATIENT HAD POOR BONE QUALITY. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20724216 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141632. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2018 · Mar 23, 2018 · Terminated
    Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page