Open Pivot™ 501DM26Recalls & reported problems
Medtronic, Inc. · MHV 501DM26 AP MITRAL · Class 3 · Implant
All models of Open Pivot (42)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169334823), received between Dec 1, 2016 and Dec 1, 2016 — classified as 1 death.
The most frequently reported specific device problem is “Mechanical Jam” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P990046); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Mechanical Jam1
Patient problems
- Mitral Regurgitation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathDec 1, 2016Mechanical Jam; Mechanical Jam
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS MECHANICAL MITRAL VALVE, THE PHYSICIAN WAS UNABLE TO ROTATE THE DEVICE INTO AN ANT-ANATOMICAL POSITION AFTER A CONSIDERABLE AMOUNT OF TIME AND EFFORT. AFTER THE HEART WAS CLOSED AND WHILE ATTEMPTING TO WEAN THE PATIENT OFF OF CARDIOPULMONARY BYPASS (CPB), A TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) SHOWED THAT ONE OF THE MITRAL VALVE LEAFLETS WAS NOT MOVING, RESULTING IN SEVERE REGURGITATION. THE PATIENT WAS RETURNED TO CPB, AND THE PHYSICIAN UNSUCCESSFULLY ATTEMPTED TO ROTATE THE DEVICE INTO A MORE OPTIMAL POSITION. THE PHYSICIAN EXPLANTED THE DEVICE, DEBRIDED THE ANNULUS OF EXCESS TISSUE, AND IMPLANTED A 24MM MECHANICAL VALVE. POST-OPERATIVELY, THE PATIENT WAS PLACED ON EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) AND TRANSFERRED TO THE INTENSIVE CARE UNIT. TWO DAYS LATER, THE PATIENT DIED.
Patient problems: Mitral Regurgitation
MAUDE report 6139699 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990046. One submission often covers many models, so check the product and lot details in each record.
- Z-1533-2020 · Feb 11, 2020 · TerminatedMedtronic Open Pivot Aortic Valve Graft, REF 502AG25, SIZE 25 mm
- Z-0634-04 · Feb 18, 2004 · TerminatedATS Open Pivot Mechanical Heart Valve
Other HEART-VALVE, MECHANICAL devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page