VALIANT® Captivia VAMC4642C150TURecalls & reported problems
Medtronic, Inc. · STENT GRAFT VAMC4642C150TU VALCAP TAPER · Class 3 · Implant
All models of VALIANT Captivia (95)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169346680), received between May 22, 2024 and May 22, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Device Damaged by Another Device1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 22, 2024Activation, Positioning or Separation Problem; Device Damaged by Another Device; Activation, Positioning or Separation Problem; Device Damaged by Another Device
VALIANT CAPTIVIA STENT GRAFTS WERE IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION. THE PROXIMAL THORACIC AORTA DIAMETER AT IMPLANTATION MEASURED 40MM, WITH THE DISSECTION SPANNING A LENGTH OF 20MM. IT WAS REPORTED DURING THE INDEX PROCEDURE, THAT THE TAPERED PORTION OF THE VALIANT CAPTIVIA (VAMC4642C150TU) STENT GRAFT WAS DELIVERED NEAR A FOLD IN THE EXISTING OPEN GRAFT REPAIR. DURING DEPLOYMENT, THE INNER CURVE OF THE DEVICE INITIALLY MADE CONTACT WITH THE INNER CURVE OF THEWALL, THE STENT GRAFT WAS THEN RELEASED AND ROTATED DISTALLY. A LONG STENT GRAFT (46X200) PIECE WAS SUCCESSFULLY PLACED AND SEAL WAS ACHIEVED. PER THE PHYSICIAN THE CAUSE OF THE DEPLOYMENT ISSUE WAS ANATOMY RELATED. IT WAS SUSPECTED THAT THE DEVICE MAY HAVE INITIALLY DEPLOYED IN CONTACT WITH THE GRAFT FOLD BEFORE RELEASE AND ROTATION. NO ADDITIONAL CLINICAL SEQUELAE WERE PROVIDED, AND THE PATIENT IS FINE.
Patient problems: Insufficient Information
MAUDE report 19371984 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page