VALIANT® Captivia VAMC3636C150TURecalls & reported problems

Medtronic, Inc. · STENT GRAFT VAMC3636C150TU VALCAP CLOSE · Class 3 · Implant

All models of VALIANT Captivia (95)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169346857), received between Nov 3, 2020 and Jun 18, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Leak/Splash1
  • Positioning Problem1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Perforation of Vessels1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2024Difficult to Remove; Positioning Problem; Difficult to Remove; Positioning Problem

    A VALIANT CAPTIVIA VAMC3636C150TU STENT GRAFT WAS INTENDED TO BE IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF A 70MM TAA . IT WAS NOTED THAT THE PATIENT HAD A PREVIOUS ASCENDING ARCH REPAIR WITH A DESCENDING THORACIC COMPONENT. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE THE FIRST VALIANT CAPTIVIA STENT GRAFT (VAMF3232C200TU SERIAL # (B)(6) WAS PLACED WITH MINIMAL RESISTANCE AND DEPLOYED SUCCESSFULLY. HOWEVER THE SECOND VALIANT CAPTIVIA STENT GRAFT (VAMC3636C150TU, SERIAL# (B)(6) WENT TO THE MID EXTERNAL ILIAC AND HAD RESISTANCE. WHEN THE PHYSICIAN PUSHED THE DEVICE THE ARTERY CLAMPED DOWN ON THE DELIVERY SYSTEM AND THE NOSE CONE AND THE OUTER GRAFT COVER GAP WAS LONGER ON THE FLUOROSCOPY IMAGE. THE PHYSICIAN ATTEMPTED TO PULL THE DEVICE BACK AND ENCOUNTERED INCREASED RESISTANCE AND CONTINUED TO TRY TO REMOVE THE DEVICE BUT WITH NO SUCCESS. THE PATIENT'S BLOOD PRESSURE STARTED FALLING AND THE PHYSICIAN IMMEDIATELY PLACED A RELIANT BALLOON AND CARRIED OUT A RETROPERITONEAL INCISION TO REPAIR THE LEFT EXTERNAL ILIAC ARTERY. THE PHYSICIAN IDENTIFIED THAT THE ARTERY WAS STUCK IN THE GAP BETWEEN THE NOSE CONE AND THE DELIVERY S YSTEM AND LOOKED LIKE IT INSIDE OUT STUCK IN THE GRAF

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information; Perforation of Vessels

    MAUDE report 19556654 ↗
  • MalfunctionNov 3, 2020Leak/Splash; Leak/Splash

    A VALIANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN 50MM FUSIFORM THORACIC AORTIC ANEURYSM. THE AORTA WAS NOTED AS SEVERELY TORTUOUS. IT WAS REPORTED APPROXIMATELY 5 YEARS POST THE INDEX PROCEDURE A DISTAL TYPE IB ENDOLEAK WAS IDENTIFIED. PER THE INVESTIGATOR THE CAUSE OF THE ENDOLEAK IS UNDETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10777154 ↗