VALIANT® Captivia VAMF4646C100TURecalls & reported problems
Medtronic, Inc. · STENT GRAFT VAMF4646C100TU VALCAP FREFL · Class 3 · Implant
All models of VALIANT Captivia (95)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169347328), received between Oct 23, 2017 and May 26, 2025 — classified as 1 death, 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Cardiac Arrest1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Vascular Dissection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 26, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT A PATIENT UNDERWENT A THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) WITH A VALIANT CAPTIVIA STENT GRAFT TO TREAT A RUPTURED PAU. DURING A FOLLOW-UP COMPUTED TOMOGRAPHY, A TYPE IA ENDOLEAK WAS IDENTIFIED, ATTRIBUTED TO THE PATIENT'S ANATOMY. TO ADDRESS THE ENDOLEAK, FIVE ENDOANCHORS WERE PLACED. THE TREATMENT RESOLVED THE TYPE IA ENDOLEAK.
Patient problems: Insufficient Information
MAUDE report 22084140 ↗DeathOct 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A VALIANT CAPTIVIA STENT GRAFT WAS IMPLANTED IN THE ENDOVASCULAR TREATMENT OF AN ACUTE DISSECTION. IT WAS REPORTED DURING THE INDEX PROCEDURE THAT FOLLOWING THE STENT GRAFT IMPLANT THAT THE PATIENT WENT INTO CARDIAC ARREST AND THE PATIENT PERFORMED COMPRESSIONS BUT THE PATIENT EXPIRED. THE PHYSICIAN REPORTED THAT THE DEATH WAS NOT ATTRIBUTED TO THE STENT GRAFT. IT WAS NOTED THAT THERE NO SALES REP PRESENT AT THE CASE DUE ITS URGENCY. AS PER THE PHYSICIAN THE CAUSE OF THE EVENT WAS DUE TO ACUTE DISSECTION AND RUPTURE OF THE DESCENDING AORTA. NO ADDITIONAL CLINICAL SEQUALAE WAS PROVIDED AND THE PATIENT HAS EXPIRED.
Patient problems: Cardiac Arrest; Vascular Dissection
MAUDE report 12583302 ↗MalfunctionOct 23, 2017Leak/Splash; Leak/Splash
A VALIANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A THORACOABDOMINAL ANEURYSM. IT WAS REPORTED THAT THE PHYSICIAN SUCCESSFULLY IMPLANTED THE VALIANT STENT GRAFT ALONG WITH SUPERIOR MESENTERIC ARTERY (SMA) AND LEFT RENAL ARTERY WITH THE SNORKEL TECHNIQUE. THE PHYSICIAN MODELLED THE VALIANT STENT GRAFT WITH BALLOONING HOWEVER THE AORTOGRAM REVEALED A TYPE IA ENDOLEAK. THE PHYSICIAN ELECTED TO SNORKEL THE CELIAC ARTERY, LENGTHEN THE SMA STENT AND THE LEFT RENAL ARTERY STENT AND IMPLANTED A 2ND VALIANT STENT GRAFT HIGHER IN THE DESCENDING THORACIC AORTA. ALL DEVICES WERE PLACED SUCCESSFULLY. BALLOONING WAS COMPLETED HOWEVER THE AORTOGRAM REVEALED THAT THE TYPE IA ENDOLEAK WAS STILL PRESENT. PER THE PHYSICIAN, THE CAUSE OF THE PROXIMAL TYPE I ENDOLEAK WAS DUE TO THE PATIENT¿S ANATOMY, THORACOABDOMINAL ANEURYSM. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT WILL BE MONITORED BY THE PHYSICIAN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6967941 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page