Quick summary

This Medtronic XPS Straightshot microdebrider is a powered surgical handpiece used in ear, nose and throat surgery. All 128 reports are malfunctions, with no injuries or deaths; most describe the handpiece overheating, and others describe blockages, vibration or noise. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
129
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 00643169350687), received between Dec 9, 2016 and Sep 28, 2026 — classified as 129 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (118 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device118
  • Mechanical Jam9
  • Use of Device Problem8
  • Vibration8
  • Noise, Audible6
  • Display or Visual Feedback Problem4
  • Intermittent Energy Output2
  • Intermittent Loss of Power2

Patient problems

  • No Clinical Signs, Symptoms or Conditions83
  • No Consequences Or Impact To Patient23
  • No Patient Involvement14
  • No Known Impact Or Consequence To Patient11

Recent adverse event reports

The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Break

    It was reported that, the M4 was not working. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911500 ↗
  • MalfunctionSep 24, 2026Overheating of Device

    It was reported that intra operatively, after the 0-degree blade was opened, the procedure could not be completed because the handpiece malfunctioned, overheated, and stopped working. The overheating was noticed when used for less than one minute. The handpiece was run at 5000 rpm speed with reciprocating rotation mode and irrigation was being used. As a result, the surgery was not finished. The hospital then used another brand of powered system with a new blade head to restart the procedure and complete the surgery. There is suspicion of connection issue with the blade. There was delay of surgery for few minutes. There was no patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50879928 ↗
  • MalfunctionAug 5, 2026Overheating of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50137304 ↗
  • MalfunctionJul 14, 2026Power Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25810841 ↗
  • MalfunctionJul 5, 2026Overheating of Device

    IT WAS REPORTED THAT PRE-OPERATION, THE M4 HANDPIECE WAS HEATING UP IN LESS THAN ONE MINUTE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25712530 ↗
  • MalfunctionDec 26, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT HANDPIECE WAS OVERHEATING. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23894627 ↗
  • MalfunctionJun 18, 2025Overheating of Device; Incorrect, Inadequate or Imprecise Result or Readings; Overheating of Device; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRIOR TO USE, THE HANDPIECE IS NOT FUNCTIONING DUE TO A BLOCKAGE INSIDE, OVERHEATING AFTER PRESSING THE FOOT SWITCH AND BLADE NOT ROTATING. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22266029 ↗
  • MalfunctionMay 26, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED DURING PRE-OP THAT THE DEVICE HAD EXCESSIVE HEAT. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22083216 ↗
  • MalfunctionApr 28, 2025Overheating of Device; Physical Resistance/Sticking; Overheating of Device; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE ROD GOT STUCK IN THE M4 MOTOR AND COULD NOT BE REMOVED. IN ADDITION, THE MOTOR GOT OVERHEATED AND THE FIRST DRILL BIT GOT MELTED. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21916542 ↗
  • MalfunctionApr 22, 2025Overheating of Device; Mechanical Jam; Overheating of Device; Mechanical Jam

    AS PER THE PRODUCT ANALYSIS, THE DEVICE PRE-REPAIR TEMPERATURE TEST RESULTS ARE GEAR: 84F, MOTOR: 86F, BEARING: 87F FOR ONE MINUTE. ALL TEMPERATURES ARE WITHIN SPECIFICATION I.E. LESS THAN 118F.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21881333 ↗
  • MalfunctionFeb 10, 2025Vibration; Vibration

    IT WAS REPORTED THAT PRIOR TO USE, THE M4 HANDPIECE HAD BUR WOBBLING. THERE WAS NO PATIENT INVOLVED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21341969 ↗
  • MalfunctionFeb 6, 2025Vibration; Vibration

    IT WAS REPORTED THAT BEFORE OPERATION, THE BUR WOBBLES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21320031 ↗
  • MalfunctionFeb 6, 2025Vibration; Vibration

    IT WAS REPORTED THAT THE BEFORE OPERATION, THE BUR WOBBLES IN M4 HANDPIECE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21319478 ↗
  • MalfunctionFeb 4, 2025Vibration; Vibration

    IT WAS REPORTED THAT DURING PRE-OP, THE BUR WAS WOBBLING. THERE WAS NO PATIENT INVOLVED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21303213 ↗
  • MalfunctionNov 13, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING PROCEDURE M4 HANDPIECE HAD EXCESSIVE HEAT. IT WAS FURTHER STATED OVERHEATING WAS NOTICED IN LESS THAN A MINUTE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20671142 ↗
  • MalfunctionOct 28, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT M4 HANDPIECE WAS OVERHEATING. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20541848 ↗
  • MalfunctionOct 21, 2024Vibration; Vibration

    IT WAS REPORTED THAT PRE OP, ABOVE 3000 RPM , IN OSCILLATION MODE DEVICE NOT SMOOTHLY WORKING WITH NEXUS. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20490629 ↗
  • MalfunctionSep 18, 2024Overheating of Device; Mechanical Jam; Intermittent Loss of Power; Overheating of Device; Mechanical Jam; Intermittent Loss of Power

    ADDITIONAL INFORMATION RECEIVED STATING DEVICE WAS OVERHEATED WHEN USED MORE THAN A MINUTE WHEN RUNNING AT 5000RPM IN OSCILLATION MODE. DEVICE DID NOT AUTOMATICALLY ACTIVATED. HANDPIECE RUN WHEN THE FOOT SWITCH PRESSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20244658 ↗
  • MalfunctionAug 22, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DEVICE DOES NOT TURN ANYMORE AND THE MOTOR GETS HOT. THERE IS NO CONSEQUENCES OR IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20024340 ↗
  • MalfunctionJul 23, 2024Overheating of Device; Overheating of Device

    ADDITIONAL INFORMATION RECEIVED STATED THAT THE HEATING WAS OBSERVED AFTER 2 MINUTES OF USE. THE HANDPIECE WAS BEING RUN A 3000 RPM IN OSCILLATE MODE WITH NO IRRIGATION BEING USED. THE PROCEDURE WAS COMPLETED MANUALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19803962 ↗
  • MalfunctionJun 24, 2024Overheating of Device; Intermittent Energy Output; Overheating of Device; Intermittent Energy Output

    ADDITIONAL INFORMATION RECEIVED STATED THAT ANOTHER UNIT WAS SET UP TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19594855 ↗
  • MalfunctionJun 3, 2024Overheating of Device; Overheating of Device

    ADDITIONAL INFORMATION RECEIVED THAT THE OVERHEATING WAS NOTICED IN LESS THAN ONE MINUTE. THE DEVICE BEING RUN AT 5000RPM AND RUN AT OSCILLATE MODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19445027 ↗
  • MalfunctionMar 22, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT INTRA OP DEVICE WAS OVERHEATING. THERE IS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18961819 ↗
  • MalfunctionMar 20, 2024Overheating of Device; Overheating of Device

    HANDPIECE WAS GETTING HOT WHEN IN USE IN 1-2 MINUTES. IT WAS RUNNING AT 5000RPM, OSCILLATION MODE, 25CC FLOWRATE. BACKUP DEVICE WAS USED FOR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18938612 ↗
  • MalfunctionFeb 12, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED, THAT DURING PRE OP. THE M4 HANDPIECE IS GETTING OVERHEAT, DURING USE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18692458 ↗

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →

Sources and updates