Skip to content

ANATOMIC PEEK PTC Cervical Fusion System 5030141Recalls & reported problems

Medtronic Sofamor Danek, Inc. · CAGE 5030141 ANATOMIC PTC 14X11X11MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169357624), received between Apr 11, 2024 and Apr 11, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Discomfort1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 11, 2024Migration; Migration

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING REVISION SURGERY. PRE-OP DIAGNOSIS WAS CERVICAL KYPHOSIS. IT WAS REPORTED THAT, PATIENT HAD DISCOMFORT DUE TO PLATE DISLODGEMENT. AFTER C5/6 DECOMPRESSION, 11MM CAGE WAS PLACED AND MIGRATION OF C5/6 CAGE ALREADY OCCURRED AFTER COMBINED POSTERIOR FIXATION. AFTER 1 WEEK POSTOPERATIVELY, FURTHER MIGRATION OF THE CAGE AND HEAD SIDE DISLODGEMENT OF THE C4-7 ZEVO PLATE OCCURRED. TOTAL REMOVAL OF PLATE AND REMOVAL OF  11MM CAGE AT C5/6, AUTOLOGOUS ILIAC BONE TRANSPLANTATION BETWEEN C5/6 VERTEBRAE WAS RETRIEVED DURING REVISION SURGERY.  THE 13MM SCREW AT C4 AND 15MM SCREW AT C7 WAS FOUND LOOSE AND THEY WERE EXPLANTED. PROCEDURE OR TECHNIQUE PERFORMED DURING INITIAL SURGERY WAS C4-7 ANTERIOR CERVICAL DISCECTOMY AND FUSION. THERE WERE NO PARTICULAR PROBLEMS WITH 5MM CAGE.  NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: Discomfort

    MAUDE report 19094240 ↗

Other Intervertebral fusion device with bone graft, cervical devices

All Intervertebral fusion device with bone graft, cervical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page