Skip to content

NC Euphora™ NCEUP3027XRecalls & reported problems

Medtronic, Inc. · BALLOON NCEUP3027X NCEUPHORA RX US · Class 2

All models of NC Euphora (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169395831), received between May 28, 2018 and Nov 11, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Deflation Problem1
  • Improper or Incorrect Procedure or Method1
  • Inflation Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2024Inflation Problem; Improper or Incorrect Procedure or Method; Material Deformation; Inflation Problem; Improper or Incorrect Procedure or Method; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE NC EUPHORA CORONARY BALLOON CATHETER TO TREAT A NON-TORTUOUS, NON-CALCIFIED LESION LOCATED WITH 70% STENOSIS IN THE MID RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS PRE-DILATED. THE DEVICE DID PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT INFLATION DIFFICULTIES WERE ENCOUNTERED DURING BALLOON INFLATION. THE BALLOON WAS INITIALLY USED, INFLATED TO NOMINAL PRESSURE, AND THEN REMOVED TO PERFORM INTRAVASCULAR ULTRASOUND (IVUS). THEN THE BALLOON WAS RE-INSERTED AND INFLATED TO 25 ATM, AND AFTER THE THIRD INFLATION THE BALLOON STARTED GROWING PROXIMALLY BEYOND THE PROXIMAL MARKER. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20651750 ↗
  • InjuryMay 28, 2018Burst Container or Vessel; Deflation Problem; Burst Container or Vessel; Deflation Problem

    AN NC EUPHORA BALLOON CATHETER WAS BEING USED TO POST DILATE A STENT THAT WAS IMPLANTED IN A MILDLY TORTUOUS, MILDLY CALCIFIED LESION WITH 95% STENOSIS IN THE DISTAL CIRCUMFLEX. THERE WAS NO ABNORMALITY IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING. THERE WERE NO ISSUES REMOVING THE DEVICE FROM THE HOOP. NO ISSUES NOTED WHEN REMOVING THE PROTECTIVE SHEATH OR REMOVING THE STYLETTE. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. IT IS INDICATED THAT THE BALLOON MAY NOT HAVE BEEN FLUSHED THROUGH THE HOOP OR PLACED IN A BOWL OF SALINE PRIOR TO USE. THE LESION WAS PRE-DILATED. THE DEVICE DID PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. SIXTY SALINE/40 CONTRAST WAS USED. NO ISSUES WERE NOTED DURING INFLATION. IT IS REPORTED THAT THE BALLOON WOULD NOT DEFLATE AT THE LESION SITE AFTER FIRST INFLATION. THE DEVICE WAS NOT MOVED OR REPOSITIONED PRIOR TO THE DEFLATION DIFFICULTIES. FIRST THE PHYSICIAN TRIED TO PULL NEGATIVE PRESSURE WITH A LARGER SYRINGE WITHOUT SUCCESS. THEN THE PHYSICIAN RE-INFLATED THE BALLOON AND PULLED NEGAT

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 7547761 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates