Skip to content

NC Euphora™ NCEUP4008XRecalls & reported problems

Medtronic, Inc. · BALLOON NCEUP4008X NCEUPHORA RX US · Class 2

All models of NC Euphora (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169396234), received between Jul 23, 2020 and Nov 15, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Material Deformation2
  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2023Burst Container or Vessel; Material Deformation; Burst Container or Vessel; Material Deformation

    AN ATTEMPT WAS MADE TO USE TWO NC EUPHORA RX BALLOON CATHETERS TO TREAT A LESION. THE DEVICES WERE INSPECTED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED. THE DEVICES PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICES. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT A BALLOON BURST OCCURRED WHEN USING BOTH DEVICES. IT WAS DETAILED THAT THE TWO NC EUPHORA BALLOON CATHETERS WERE USED TO POST DILATE A STENT. THE 4.00X8MM NC EUPHORA WAS INFLATED TWICE AT 16 ATM FOR 10SECONDS. THE BALLOON RUPTURED DURING THE SECOND DEFLATION. THE 4.0X20MM NC EUPHORA WAS THEN DILATED WITH TWO INFLATIONS AT 14ATM FOR 15SECONDS AND A THIRD INFLATION AT 14ATM FOR 10SECONDS. THE BALLOON RUPTURE OCCURRED DURING THE THIRD DEFLATION. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18139854 ↗
  • MalfunctionNov 26, 2021Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    AN ATTEMPT WAS MADE TO USE A RESOLUTE ONYX DRUG ELUTING STENT AND AN NC EUPHORA BALLOON CATHETER TO TREAT A LIMITED TORTUOUS LESION WITH 85% STENOSIS IN THE MID LAD. THE DEVICES WERE INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICES DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICES. EXCESSIVE FORCE WAS NOT USED. THE DEVICES WERE NOT KINKED AND RE-STRAIGHTENED DURING USE. IT WAS REPORTED THAT A STENT DISLODGEMENT OCCURRED AT THE CATHETER DURING ADVANCEMENT THROUGH THE NON-MEDTRONIC (MDT) GUIDE CATHETER. IT WAS ALSO REPORTED THAT THE NC EUPHORA CATHETER BECAME DETACHED IN TWO PIECES IN THE GUIDE CATHETER. IT WAS SUGGESTED THAT THE BALLOON MAY HAVE BECOME TWISTED DUE TO TWO NON-MDT GUIDE WIRES INSIDE THE GUIDE CATHETER. IT WAS ATTEMPTED TO PASS THE RESOLUTE ONYX THOUGH AND WAS ALSO SUGGESTED THAT THE STENT BECAME PARTIALLY DISLODGED OR GOT CAUGHT WITH THE BALLOON. THE DEVICES WERE EASILY REMOVED. THE DEVICES NEVER REACHED THE PATIENT VESSEL. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12877341 ↗
  • MalfunctionJul 23, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE ONE NC EUPHORA RX PTCA BALLOON CATHETER TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED LESION LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT THE DEVICE DETACHED AT THE CATHETER DURING ADVANCEMENT THROUGH THE GUIDE CATHETER AND NEVER EXITED THE GUIDE. THE DEVICE WAS NOT KINKED OR RESTRAIGHTENED DURING USE. THE DETACHED PORTION WAS REMOVED BY PULLING THE DEVICE OUT OF THE GUIDE. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10312637 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page